HOL ML 5MM ENDO APPLIER
Report
- Report Number
- 3011137372-2022-00022
- Event Type
- Malfunction
- Date Received
- February 3, 2022
- Date of Event
- January 26, 2022
- Report Date
- January 26, 2022
- Manufacturer
- TELEFLEX MEDICAL
- Product Code
- GDO
- UDI-DI
- 24026704696813
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
(B)(4). THE DEVICE HAS NOT BEEN RETURNED FOR INVESTIGATION. TELEFLEX WILL CONTINUE TO MONITOR AND TREND RELATED EVENTS.
(B)(6) THE DHR FOR THE RETURNED INSTRUMENT WAS REVIEWED AND FOUND COMPLETELY WITHOUT ANY IRREGULARITIES. THIS INSTRUMENT WAS PRODUCED AT THE TECOMET, INC. KENOSHA WI FACILITY AS PART OF A 50 PC. LOT IN JUNE OF 2020. EVALUATION OF THE RETURNED INSTRUMENT SHOWS THAT THE TIPS ARE LOOSE AND MISALIGNED, AND THE JAW PIVOT PIN IS SLIGHTLY STICKING OUT OF ONE SIDE OF THE OUTER TUBE ASSEMBLY AND PARTIALLY SUNKEN INTO THE OTHER SIDE. WE ARE ABLE TO VALIDATE THIS COMPLAINT. AFTER THE INITIAL EVALUATION THIS INSTRUMENT WAS DIS-ASSEMBLED IN ORDER TO EVALUATE ITS INTERNAL COMPONENTS AND IT WAS FOUND THAT THE INNER DRIVE ROD (N00185) BOSSES ARE BOTH DAMAGED WHERE THEY ENGAGE THE JAWS. WE SUSPECT THAT THE DAMAGED DRIVE ROD BOSSES CAUSED THE JAWS TO BECOME SLIGHTLY LOOSE AND MISALIGNED IN THE CLOSED POSITION. WE ARE UNABLE TO DETERMINE WHAT CAUSED THE DRIVE ROD BOSSES TO BECOME DAMAGED AND FOR THE JAW PIVOT PIN TO BE SUNKEN IN ON ONE SIDE OF THE OUTER TUBE ASSEMBLY AND SLIGHTLY STICKING OUT THE OTHER SIDE BUT MISHANDLING OF THIS DEVICE AT THE END USER'S FACILITY IS SUSPECTED. ALL 50 INSTRUMENTS FROM THIS LOT WERE 100% VISUALLY INSPECTED AND FUNCTION TESTED PRIOR TO SHIPMENT TO THE CUSTOMER AS THIS IS A STANDARDIZED PROCEDURE AT THIS FACILITY FOR THIS PRODUCT LINE. DUE TO THESE FINDINGS, NO FURTHER ACTIONS WILL BE TAKEN IN RESPONSE TO THIS COMPLAINT AND THIS RECORD WILL BE DEEMED CLOSED.
THE PIVOT PIN WAS FOUND LOOSE DURING A PRETEST BEFORE USE. THEREFORE, A NEW UNIT WAS USED INSTEAD. THE APPLIER WAS PURCHASED BY THE HOSPITAL WITHIN 1 YEAR.
THE PIVOT PIN WAS FOUND LOOSE DURING A PRETEST BEFORE USE. THEREFORE, A NEW UNIT WAS USED INSTEAD. THE APPLIER WAS PURCHASED BY THE HOSPITAL WITHIN 1 YEAR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1414629 | HOL ML 5MM ENDO APPLIER | APPLIER, SURGICAL, CLIP | GDO | TELEFLEX MEDICAL | IPN915189 | 06B2098688 | 24026704696813 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |