FDA Adverse Event Malfunction Summary report: N

CANNULATED STRAIGHT DRIVER

MDR report key: 13440597 · Received February 3, 2022

Report

Report Number
0001822565-2022-00178
Event Type
Malfunction
Date Received
February 3, 2022
Date of Event
January 18, 2022
Report Date
May 2, 2022
Manufacturer
ZIMMER BIOMET, INC.
Product Code
KWT
PMA / PMN Number
K103404
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4). DEVICE HAS BEEN RECEIVED BY ZIMMER BIOMET FOR INVESTIGATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. VISUAL EXAMINATION OF THE RETURNED PRODUCT VERIFIED ITEM AND LOT NUMBER. THE CANNULATED STRAIGHT DRIVER EXHIBITS SIGNS OF REPEATED USE (NICKED OR GOUGED) BUT IS NOT FRACTURED. PERFORMED DIMENSIONAL ANALYSIS ON CANNULATED STRAIGHT DRIVER LOT 60698536 REFERENCING NOTE#1 ON PRINT. IT'S CONFIRMED THE FUNCTIONAL GAGE ITEM# 25-1002-511-00 PASSES AND FUNCTIONS SMOOTHLY. REVIEW OF THE DEVICE HISTORY RECORD(S) IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING. NO PROBLEM FOUND FOR PART 00430707400; RETURNED PRODUCT FOUND TO BE CONFORMING TO SPECIFICATIONS. UPON REASSESSMENT OF THE REPORTED EVENT, IT WAS DETERMINED TO BE NOT REPORTABLE. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT A CANNULATED STRAIGHT DRIVER WAS BROKEN DURING A PROCEDURE. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1749607 CANNULATED STRAIGHT DRIVER ORTHOPAEDIC PROSTHESIS IMPLANTATION INSTRUMENT, REUSABLE KWT ZIMMER BIOMET, INC. NI 60698536

Patients

Seq Age Sex Outcome Treatment
1 Male