FDA Adverse Event Death Summary report: N

ACIST

MDR report key: 1343826 · Received March 12, 2009

Report

Report Number
2134243-2009-00001
Event Type
Death
Date Received
March 12, 2009
Date of Event
January 6, 2009
Report Date
March 10, 2009
Manufacturer
ACIST MEDICAL SYSTEMS, INC.
Product Code
DXT
PMA / PMN Number
K991103
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ACIST'S MEDICAL ADVISORY BOARD MEMBER CONTACTED THE HOSPITAL'S CHARGE TECHNOLOGIST - RADIOLOGY AND DISCUSSED THE EVENT. PER THIS DISCUSSION, ACIST'S MEDICAL ADVISORY BOARD MEMBER STATE THAT THE ROOT CAUSE OF THE EVENT IS CATHETER-RELATED DISSECTION. ALTHOUGH THERE IS NO EVIDENCE OF MALFUNCTION OF THE ACIST ANGIOGRAPHIC CONTRAST INJECTION SYSTEM, MODEL CVI, THE INJECTION SYSTEM IS REQUESTED TO BE RETURNED TO ACIST FOR TESTING, PER ACIST'S STANDARD PROCEDURE. THE CVI INJECTION SYSTEM HAS BEEN IN USE AT THE HOSPITAL SINCE THE EVENT OCCURRED. THE ACIST STERILE DISPOSABLE KITS WERE REQUESTED TO BE RETURNED, BUT DUE TO THE ONE-MONTH LAPSE BETWEEN THE EVENT AND THE REPORT TO ACIST, THESE HAD ALREADY BEEN DISCARDED BY THE HOSPITAL.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT DURING A DIAGNOSTIC CARDIAC CATHETERIZATION PROCEDURE, USING THE ACIST ANGIOGRAPHIC CONTRAST INJECTION SYSTEM, MODEL CVI AND A CORDIS EBA 3.5 CATHETER, A DISSECTION OF A PATIENT'S CORONARY ARTERY (LEFT MAIN AND CIRCUMFLEX) OCCURRED (DATE OF EVENT 2009). THE PATIENT EXPERIENCED CHEST PAIN AND ST SEGMENT ELEVATION. SURGERY WAS PERFORMED ON THE PATIENT TO REPAIR THE DISSECTION. THE PATIENT DIED IN 2006. THE HOSPITAL'S CHARGE TECHNOLOGIST - RADIOLOGY REPORTED THAT THE CAUSE OF THE EVENT WAS THE EBA 3.5 CATHETER PLACED DEEPLY INTO THE LEFT MAIN CORONARY ARTERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACIST ANGIOGRAPHIC INJECTION SYSTEM DXT ACIST MEDICAL SYSTEMS, INC. CVI

Patients

Seq Age Sex Outcome Treatment
1 51 YR Death 3.5 CATHETER:| CORDIS EBA 3.5 CATHETER:| NITROGLYCERINE:| CORDIS JUDKINS JL 4:| FENTANYL: