FDA Adverse Event Malfunction Summary report: N

HIGH DEFINITION LCD MONITOR

MDR report key: 13435247 · Received February 3, 2022

Report

Report Number
8010047-2022-02345
Event Type
Malfunction
Date Received
February 3, 2022
Date of Event
January 5, 2022
Report Date
April 25, 2022
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
FET
UDI-DI
04953170363498
PMA / PMN Number
K102379
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VM
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE EVALUATION FOUND THE MONITOR TURNS OFF POWER AUTOMATICALLY AFTER TURNING ON SOME MINUTES DUE TO A DEFECTIVE POWER SUPPLY BOARD. THE INVESTIGATION IS ONGOING; THEREFORE, THE ROOT CAUSE OF THE REPORTED ISSUE CANNOT BE DETERMINED AT THIS TIME. HOWEVER, IF ADDITIONAL INFORMATION BECOMES AVAILABLE A FOLLOW UP MEDICAL DEVICE REPORT WILL BE SUPPLEMENTED ACCORDINGLY.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S FINAL INVESTIGATION A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. BASED ON THE RESULTS OF THE LEGAL MANUFACTURER'S INVESTIGATION, THE CAUSE OF THE DEFECTIVE POWER SUPPLY BOARD COULD NOT BE IDENTIFIED. NO DEFECT INSIDE THE DEVICE OTHER THAN THE PHENOMENON POINTED OUT. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

THE OLYMPUS (OVN) TECHNICAL SERVICE ENGINEER BECAME AWARE THAT THE ASSET HIGH DEFINITION LIQUID CRYSTAL DISPLAY MONITOR ¿DID NOT POWER UP WHEN INSPECTED BEFORE DELIVERING TO DEMONSTRATION." NO HARM WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1812786 HIGH DEFINITION LCD MONITOR HIGH DEFINITION LCD MONITOR FET OLYMPUS MEDICAL SYSTEMS CORP. OEV262H 04953170363498

Patients

Seq Age Sex Outcome Treatment
1 Unknown