SYNCHRON CX9 ALX
Report
- Report Number
- 2050012-2009-00035
- Event Type
- Malfunction
- Date Received
- March 13, 2009
- Date of Event
- February 25, 2009
- Report Date
- March 13, 2009
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJC
- PMA / PMN Number
- K950958
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
QC WAS WITHIN LAB ESTABLISHED RANGES. NO OTHER QUESTIONABLE CRE3 RESULTS WERE FOUND. A FIELD SERVICE ENGINEER WAS DISPATCHED: THE FSE PERFORMED MAINTENANCE ON THE CUP AND VERIFIED PERFORMANCE. AS OF 2009, NO FURTHER EVENTS HAVE BEEN REPORTED. A CLEAR ROOT CAUSE FOR THIS EVENT HAS NOT BEEN DETERMINED. A MALFUNCTION WILL BE ASSUMED FOR THE PURPOSE OF THIS REPORT.
A CUSTOMER TESTED PATIENT SAMPLE FOR CREATININE (CRE3) AND A RESULT OF 0.9MG/DL WAS OBTAINED. THE RESULT WAS REPORTED OUT OF THE LAB AND QUESTIONED BY THE PHYSICIAN. THE ORIGINAL SAMPLE WAS RE-TESTED AND REPEATED CRE3 RESULT WAS 1.4MG/DL. IN ADDITION, THE ORIGINAL SAMPLE WAS TESTED ON A DIFFERENT INSTRUMENT IN THE CUSTOMER'S LAB AND CRE3 RESULT OF 1.5MG/DL WAS OBTAINED. THE CRE3 RESULT WAS AMENDED. 6. IT IS UNKNOWN IF PATIENT TREATMENT WAS INITIATED OR WITHHELD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON CX9 ALX | CLINICAL CHEMISTRY ANALYZER | JJC | BECKMAN COULTER INC. | CX9 ALX |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |