FDA Adverse Event Malfunction Summary report: N

BD ULTRA-FINE¿ PRO PEN NEEDLE

MDR report key: 13427821 · Received February 2, 2022

Report

Report Number
9616656-2022-00081
Event Type
Malfunction
Date Received
February 2, 2022
Date of Event
January 18, 2022
Report Date
February 24, 2022
Manufacturer
BECTON DICKINSON AND CO.
Product Code
FMI
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL YES, D10: RETURNED TO MANUFACTURER ON: INVESTIGATION SUMMARY: THREE OPEN 32G X 4MM PEN NEEDLE SAMPLES AND ONE PHOTO WERE RETURNED FROM LOT. NO. 1082141, CAT. NO. 320561. VISUAL EXAMINATION WAS CARRIED OUT ON THE RETURNED SAMPLES AND PHOTO AND A BROKEN NON PATIENT END OF CANNULA WAS OBSERVED ON TWO SAMPLES AND NO ISSUES WERE OBSERVED WITH THE REMAINING SAMPLE. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED AS PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. AS ALL SAMPLES RETURNED WERE OPEN IT IS NOT POSSIBLE TO CONFIRM THIS DEFECT TO BE MANUFACTURING RELATED. H3 OTHER TEXT : SEE H10.

Additional Manufacturer Narrative · 0

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD ULTRA-FINE¿ PRO PEN NEEDLE WAS DAMAGED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM GERMAN: "CUSTOMER CLAIMS THAT THE NEEDLES ARE DAMAGED OFTEN."

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD ULTRA-FINE¿ PRO PEN NEEDLE WAS DAMAGED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6): "CUSTOMER CLAIMS THAT THE NEEDLES ARE DAMAGED OFTEN."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
85074 BD ULTRA-FINE¿ PRO PEN NEEDLE HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON DICKINSON AND CO. 1082141

Patients

Seq Age Sex Outcome Treatment
1 Unknown