FDA Adverse Event Malfunction Summary report: N

ALINITY C PROCESSING MODULE

MDR report key: 13425946 · Received February 2, 2022

Report

Report Number
3016438761-2022-00050
Event Type
Malfunction
Date Received
February 2, 2022
Date of Event
January 11, 2022
Report Date
March 18, 2022
Manufacturer
ABBOTT LABORATORIES
Product Code
JJE
UDI-DI
00380740137380
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. PATIENT IDENTIFIER SIDS: (B)(6).

Additional Manufacturer Narrative · 0

ABBOTT FIELD SERVICE (FS) INVESTIGATED THE ISSUE ON SITE. FS TROUBLESHOOTING INCLUDED CORRECTING THE WASH HEAD TUBING WHICH WAS INCORRECTLY INSTALLED BY THE CUSTOMER. THIS RESOLVED THE ISSUE. THERE HAVE BEEN NO FURTHER REPORTS OF DISCREPANT RESULTS SINCE THE TUBING WAS REINSTALLED. A REVIEW OF THE ALINITY C SN# (B)(6) SERVICE HISTORY WAS NOT ABLE TO IDENTIFY OR CONFIRM ANY ADDITIONAL LIKELY CAUSES. A REVIEW OF HISTORICAL DATA REVEALED NO TRENDS, SYSTEMIC ISSUES, OR RELATED NONCONFORMANCES. A REVIEW OF THE CLINICAL CHEMISTRY SYSTEMS REVEALED FOUND NO SYSTEMIC ISSUES OR TRENDS FOR THE ISSUE ASSOCIATED WITH THIS TICKET (ERRATIC/DISCREPANT RESULTS). A REVIEW OF LABELLING ADEQUATELY ADDRESSES SAMPLE RESULTS OBSERVED PROBLEMS FOR ERRATIC / DISCREPANT RESULTS. BASED ON THE INVESTIGATION NO SYSTEMIC ISSUE OR PRODUCT DEFICIENCY WAS IDENTIFIED FOR THE ALINITY C ANALYZER SN# (B)(6).

Description of Event or Problem · 0

THE CUSTOMER REPORTED FALSELY DECREASED SODIUM AND CALCIUM RESULTS GENERATED ON THE ALINITY C ANALYZER ON MULTIPLE PATIENTS ON (B)(6) 2022. RESULTS PROVIDED: SID (B)(6): SODIUM = 100/ 145 / 143 MMOL/L, SID (B)(6): CALCIUM = 4.5 / 10.5 / 10.6 MG/DL, SID (B)(6): CALCIUM = 3.0 / 9.3 / 9.3 MG/DL, SID (B)(6): CALCIUM = 5.4 / 9.8 / 9.8 MG/DL, SID (B)(6): SODIUM = 114 / 139 / 143 MMOL/L, SID (B)(6): SODIUM = 118 / 143 MMOL/L; CALCIUM = 4.7 / 9.9 MG/DL, SID (B)(6): CALCIUM = 3.9 / 9.8 MG/DL. REFERENCE RANGES: SODIUM: 136-145 MMOL/L; CALCIUM: 8.4-10.2 MG/DL NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
71851 ALINITY C PROCESSING MODULE ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE ABBOTT LABORATORIES 03R6701 00380740137380

Patients

Seq Age Sex Outcome Treatment
1 Unknown ALNTY C CALCIUM RGT 400, 07P57-20, 49044UN21| ALNTY C CALCIUM RGT 400, 07P57-20, 49044UN21| ALNTY C ICT SAMPLE DILU, 07P53-20, 02358UN21| ALNTY C ICT SAMPLE DILU, 07P53-20, 02358UN21