FDA Adverse Event Malfunction Summary report: N

VELA VENTILATOR

MDR report key: 13425751 · Received February 2, 2022

Report

Report Number
2021710-2022-15436
Event Type
Malfunction
Date Received
February 2, 2022
Date of Event
January 5, 2022
Report Date
January 5, 2022
Manufacturer
VYAIRE MEDICAL
Product Code
CBK
UDI-DI
10846446001419
PMA / PMN Number
K093094
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H11-ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER REVEALED THAT THE REPORTED ISSUE OCCURRED DURING PREVENTIVE MAINTENANCE. PLEASE DISREGARD AND VOID VYAIRE REFERENCE NUMBER:(B)(4) UNDER MEDWATCH REPORT NAME:MWR-20012022-0000011930.

Additional Manufacturer Narrative · 0

(B)(4). A VYAIRE FIELD SERVICE REPRESENTATIVE(FSR) EVALUATED THE DEVICE ONSITE AND WAS ABLE TO VERIFY THAT THE VALUES ARE MUCH LOWER THAN THE 20% +/- FOR VTE (END-TIDAL VOLUME). THE EXHALATION PRESSURE TRANSDUCER WAS CALIBRATED,PASSED AND ACCEPTED BY THE UNIT BUT IT DID NOT RESOLVE THE LOW VTE ISSUE. THE EXHALATION FLOW TRANSDUCER WAS ALSO CALIBRATED BUT CANNOT GET A READING ON THE CITREX H5 TO APPLY 3CMH20. WHEN APPLYING PRESSURE, THE CITREX VALUES FOR CMH20 TO NOT MOVE, STAYS AT ZERO. UPON FURTHER INSPECTION, THE CONNECTOR ASSEMBLY FOR THE FLOW SENSOR WAS MISSING THEIR O-RINGS, DESPITE THE UNIT PASSING THE LEAK TEST. THE UNIT WAS REPLACED,OVP (OPERATIONAL VERIFICATION PROCEDURE) WAS PERFORMED AND VERIFIED THAT ALL VALUES ARE WITHIN SPECIFICATIONS. VYAIRE MEDICAL WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 0

THE CUSTOMER REPORTED TO VYAIRE MEDICAL THAT THE VOLUMES ARE VERY LOW ON THE VELA VENTILATOR. IT WAS SET @500 BUT READING IN THE 200'S. AT THIS TIME, THERE IS NO INFORMATION REGARDING PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
60481 VELA VENTILATOR VENTILATOR, CONTINUOUS, FACILITY USE CBK VYAIRE MEDICAL VELA 10846446001419

Patients

Seq Age Sex Outcome Treatment
1 Unknown