FDA Adverse Event Malfunction Summary report: N

UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM

MDR report key: 1342230 · Received January 23, 2009

Report

Report Number
2122870-2009-00020
Event Type
Malfunction
Date Received
January 23, 2009
Date of Event
January 12, 2009
Report Date
January 23, 2009
Manufacturer
BECKMAN COULTER, INC.
Product Code
MMI
PMA / PMN Number
K060256
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NO INFORMATION

Narratives

Additional Manufacturer Narrative · 1

THE SAMPLE WAS COLLECTED IN A SODIUM HEPARIN TUBE AND IT WAS CENTRIFUGED AT AN UNSPECIFIED SPEED FOR 10 MINS. BASED ON AVAILABLE INFO, THE CUSTOMER CALIBRATED NEW ACCU TNI REGENT IN EARLY 2009 THREE TIMES BEFORE OBTAINING ACCEPTABLE RESULTS. QC CHARTS SHOW THE CUSTOMER'S QC SHIFTED HIGH IMMEDIATELY UPON SWITCHING TO THE NEW REAGENT. THE CUSTOMER PERFORMED A SYSTEM CHECK EIGHT DAYS LATER, WHICH FAILED DUE TO QUANTITY NOT SUFFICIENT (QNS) FAGS. THE CUSTOMER DID NOT REPEAT THE SYSTEM CHECK BUT COMMONLY THIS TYPE OF FAILURE IS DUE TO NOT ENOUGH SYSTEM CHECK SOLUTION IN THE SAMPLE CUP. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED: THE FSE PERFORMED A PREVENTIVE MAINTENANCE (PM) TO THE INSTRUMENT. THE FSE VERIFIED THE REPAIR PER ESTABLISHED PROCEDURES AND RESULTS MET PUBLISHED PERFORMANCE SPECIFICATIONS. A CLEAR ROOT CAUSE HAS NOT BEEN DETERMINED TO DATE FOR THIS EVENT. A MALFUNCTION WILL BE ASSUMED FOR THE PURPOSE OF THIS REPORT.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING AN ERRONEOUSLY ELEVATED TROPONIN (ACCU TNI) RESULT THAT WAS GENERATED BY THE UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM FOR ONE PT. A PT SAMPLE WAS TESTED FOR ACCU TNI AND A RESULT OF 0.54NG/ML WAS OBTAINED. THE ORIGINAL SAMPLE WAS RE-TESTED SEVERAL TIMES AND REPEATED RESULTS WERE IN THE RANGE OF 0.02-0.04 NG/ML. THE RESULTS WERE NOT REPORTED OUT OF THE LAB. IT IS UNK IF PT TREATMENT WAS AFFECTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER MMI BECKMAN COULTER, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA