COBAS® SARS-COV-2 - 480T
Report
- Report Number
- 2243471-2022-00154
- Event Type
- Malfunction
- Date Received
- February 1, 2022
- Date of Event
- January 3, 2022
- Report Date
- May 23, 2022
- Manufacturer
- ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
- Product Code
- QJR
- PMA / PMN Number
- EUA200009
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
AN INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION COMPLETES. THE CUSTOMER ISSUE HAS BEEN ALLEGED ON THE COBAS 6800 SYSTEM (PRODUCT CATALOG NUMBER 05524245001; UDI: (B)(4); PMA 510K: BK140196; PRODUCT CODE: MZA). THE TEST USED ON THE COBAS 6800 SYSTEM IS THE COBAS SARS-COV-2 FOR USE ON THE COBAS 6800-8800 SYSTEM (EUA(B)(4), PRODUCT CODE: QJR). THE PRODUCT CATALOG NUMBER FOR THE TEST (480T) IS 09343733190 AND THE UDI IS (B)(4). (B)(4).
THE ANALYTICAL UNIT OF THE COBAS ANALYZER WAS RETURNED AND AN INVESTIGATION ON THE RETURNED UNIT FOUND THE UNIT TO BE WITHIN SPECIFICATIONS. A CAPA WAS INITIATED. PER THE INVESTIGATION, THE UNDERLYING TECHNICAL ROOT CAUSE IS OPTICAL SPATIAL CROSS TALK, WHICH IN SPECIFIC SITUATIONS (HIGH POSITIVITY RATE ON THE PLATE WITH SIMILAR CT VALUES OF TRUE POSITIVES WITH HIGH RFI VALUES AROUND A NEGATIVE SAMPLE) IN COMBINATION WITH THE ASAP (ASSAY SPECIFIC ANALYSIS PACKAGE) SETTING MAY LEAD TO FALSE POSITIVES. ASAP IS A SOFTWARE WHICH CONTAINS ALL ASSAY SPECIFIC PARAMETERS AND DEFINITIONS. BASED ON THE EVALUATION OF GLOBAL COMPLAINT DATA AND THE QUANTITY OF TESTS SOLD GLOBALLY, THE FREQUENCY OF OCCURRENCE IS ESTIMATED TO BE (B)(4). A FALSE POSITIVE RATE ANYWHERE IN THIS RANGE WOULD NOT CAUSE AN IMPACT TO THE OVERALL PERFORMANCE OF THE TEST OR THE TEST¿S SPECIFICITY. WHILE A FALSE-POSITIVE RESULT FOR SARS-COV-2 COULD LEAD TO COHORTING WITH OTHERS WITH TRUE SARS-COV-2 INFECTION AND TRANSMISSION OF THE INFECTION TO THE ERRONEOUSLY DIAGNOSED PATIENT (INCLUDING THOSE AT HIGH RISK DUE TO COMORBIDITIES), PRIOR VACCINATION AND THE AVAILABILITY OF ANTIVIRAL MEDICATIONS AND OTHER TREATMENT MODALITIES CAN REDUCE THE LIKELIHOOD SEVERE ILLNESS, HOSPITALIZATION AND/OR DEATH. THE LATTER ALSO MITIGATES THE RISK FOR THOSE WHO MAY DEFER PROTECTIVE MEASURES BECAUSE OF ASSUMED NATURAL IMMUNITY. HOWEVER, PATIENTS WITH RISK FACTORS FOR SEVERE DISEASE WOULD BE LESS LIKELY TO RELAX THEIR PRECAUTIONS. TAKING THE LOW RATE OF FALSE POSITIVE OCCURRENCE AND THE LOW PROBABILITY OF SUBSEQUENT PATIENT SAFETY IMPACTS TOGETHER, THE FALSE POSITIVES ARE UNLIKELY TO CAUSE ADVERSE TRANSIENT OR SEVERE ADVERSE EVENTS. ADDITIONALLY, NO HARM WAS INDICATED IN THE CURRENT CASE. AN UPDATED ASAP VERSION WITH OPTIMIZED SETTINGS WILL BE MADE AVAILABLE. (B)(4).
IN LIGHT OF THE COVID-19 PANDEMIC AND THE SUBSEQUENT EMERGENCY USE AUTHORIZATIONS (EUAS) FOR SARS-COV-2 DIAGNOSTIC TESTS, THE AGENCY HAS REQUESTED HEIGHTENED REPORTING BEYOND THE REASONABLY SUGGESTS REQUIREMENTS OF 803 TO INCLUDE ALLEGATIONS OF FALSE POSITIVE OR FALSE NEGATIVE RESULTS INDEPENDENT OF HARM OR MALFUNCTION OR OFF-LABEL USE. PURSUANT TO THE AGENCY¿S INSTRUCTION, WE HEREBY SUBMIT THIS MDR. A CUSTOMER FROM AUSTRALIA SUSPECTED VARIOUS FALSE POSITIVE SARS-COV-2 RESULTS WHEN USING THE COBAS SARS-COV-2 TEST ON THE COBAS 6800 SYSTEM (SN# (B)(4)). SPECIFICALLY, 2 PATIENT SAMPLES GENERATED A TARGET 1 POSITIVE AND TARGET 2 NEGATIVE RESULT IN THE INITIAL TEST RUN. RETEST OF THE SAMPLES ON A COBAS 8800 SYSTEM (SN# (B)(4)) GENERATED A NEGATIVE RESULT FOR BOTH TARGET 1 AND TARGET 2. THE INITIAL POSITIVE RESULTS WERE REPORTED OUT. THESE RESULTS HAVE NOT BEEN QUESTIONED BY THE CLINICIAN OR PATIENT AND NO HARM HAS BEEN ALLEGED AS A RESULT OF THESE FALSE POSITIVES. AN INVESTIGATION IS ONGOING. TWO MDRS, ONE PER EACH RESULT, WILL BE FILED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 975368 | COBAS® SARS-COV-2 - 480T | REAGENTS, 2019-NOVEL CORONAVIRUS NUCLEIC ACID | QJR | ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG | NA | H07603 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |