BD VACUTAINER® K2 EDTA (K2E) 3.6MG BLOOD COLLECTION TUBES
Report
- Report Number
- 1917413-2022-00038
- Event Type
- Malfunction
- Date Received
- February 1, 2022
- Date of Event
- December 30, 2021
- Report Date
- April 5, 2022
- Manufacturer
- BECTON, DICKINSON & CO. (BROKEN BOW)
- Product Code
- JKA
- UDI-DI
- 50382903678414
- PMA / PMN Number
- BK050036
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
"MULTIPLE LOT NUMBERS: THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 1014077. MEDICAL DEVICE EXPIRATION DATE:2022-04-30. DEVICE MANUFACTURE DATE: 2021-01-14. MEDICAL DEVICE LOT #: 1040927. MEDICAL DEVICE EXPIRATION DATE: 2022-05-31. DEVICE MANUFACTURE DATE: 2021-02-09. MEDICAL DEVICE LOT #: 1167580. MEDICAL DEVICE EXPIRATION DATE: 2022-09-30. DEVICE MANUFACTURE DATE: 2021-06-16. MEDICAL DEVICE LOT #: 1228046 . MEDICAL DEVICE EXPIRATION DATE: 2022-11-30. DEVICE MANUFACTURE DATE: 2021-08-16. MEDICAL DEVICE LOT #: 1014077. MEDICAL DEVICE EXPIRATION DATE: 2022-04-30. DEVICE MANUFACTURE DATE: 2021-01-14. MEDICAL DEVICE LOT #: 1040927. MEDICAL DEVICE EXPIRATION DATE: 2022-05-31. DEVICE MANUFACTURE DATE: 2021-02-09. MEDICAL DEVICE LOT #: 1167580 . MEDICAL DEVICE EXPIRATION DATE: 2022-09-30. DEVICE MANUFACTURE DATE: 2021-06-16. MEDICAL DEVICE LOT #: 1228046. MEDICAL DEVICE EXPIRATION DATE: 2022-11-30. DEVICE MANUFACTURE DATE: 2021-08-16. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H.6. INVESTIGATION: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR EVALUATION. THEREFORE, RETENTION SAMPLES OF THE INCIDENT LOT WERE SELECTED FROM BD INVENTORY FOR EVALUATION AND UPON COMPLETION, NO ISSUES RELATING TO CLOTTING WERE OBSERVED. 5 PRODUCTION LOT IN-HOUSE RETENTION SAMPLES FROM EACH LOT AFFECTED WERE VISUALLY INSPECTED WITH NO ISSUES BEING IDENTIFIED AND SUBMITTED TO THE CHEMISTRY LAB FOR TESTING. ALL RESULTS WERE WITHIN THE SPECIFICATION LIMITS OF 2.67MG ¿ 4.84MG FOR CATALOG 367841. NO DIFFICULTIES WERE ENCOUNTERED DURING BLOOD COLLECTION AND ALL TUBES APPEARED TO EXHIBIT PROPER FILL. BD WAS UNABLE TO CONFIRM THE CUSTOMER¿S INDICATED FAILURE (CLOTTING) BECAUSE THE DEFECT WAS NOT EVIDENT IN THE TESTING OF THE COMPLAINT LOT SAMPLES. LABORATORY ANALYSIS OF RETAIN AND CONTROL SAMPLES DEMONSTRATED CLINICALLY ACCEPTABLE PERFORMANCE FOR ALL VISUAL OBSERVATIONS EVALUATED. ALL EDTA TUBES PERFORMED AS EXPECTED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE CLOTTING. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE.
"MULTIPLE LOT NUMBERS: THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 1014077. MEDICAL DEVICE EXPIRATION DATE:2022-04-30. DEVICE MANUFACTURE DATE: 2021-01-14. MEDICAL DEVICE LOT #: 1040927. MEDICAL DEVICE EXPIRATION DATE: 2022-05-31. DEVICE MANUFACTURE DATE: 2021-02-09. MEDICAL DEVICE LOT #: 1167580. MEDICAL DEVICE EXPIRATION DATE: 2022-09-30. DEVICE MANUFACTURE DATE: 2021-06-16. MEDICAL DEVICE LOT #: 1228046 . MEDICAL DEVICE EXPIRATION DATE: 2022-11-30. DEVICE MANUFACTURE DATE: 2021-08-16. MEDICAL DEVICE LOT #: 1014077. MEDICAL DEVICE EXPIRATION DATE: 2022-04-30. DEVICE MANUFACTURE DATE: 2021-01-14. MEDICAL DEVICE LOT #: 1040927. MEDICAL DEVICE EXPIRATION DATE: 2022-05-31. DEVICE MANUFACTURE DATE: 2021-02-09. MEDICAL DEVICE LOT #: 1167580 . MEDICAL DEVICE EXPIRATION DATE: 2022-09-30. DEVICE MANUFACTURE DATE: 2021-06-16. MEDICAL DEVICE LOT #: 1228046. MEDICAL DEVICE EXPIRATION DATE: 2022-11-30. DEVICE MANUFACTURE DATE: 2021-08-16. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED WHEN USING THE BD VACUTAINER® K2 EDTA (K2E) 3.6MG BLOOD COLLECTION TUBES, THE DEVICE EXPERIENCED CLOTTING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: IT WAS REPORTED THAT THE TUBES ARE CLOTTING. WE PROVIDE SUPPLIES AND SERVICES FOR CLINICAL TRIALS RESEARCH. WE WERE NOTIFIED BY ONE OF THE CLINICAL SITES THAT SOME OF CATALOGUE NUMBER 367841 - 2ML K2EDTA BLOOD COLLECTION TUBES WERE CLOTTED. WOULD YOU PLEASE CONFIRM THE EXPECTED STABILITY WINDOW OF WHOLE BLOOD IN THIS BCT AT AMBIENT TEMPERATURE?
IT WAS REPORTED WHEN USING THE BD VACUTAINER® K2 EDTA (K2E) 3.6MG BLOOD COLLECTION TUBES, THE DEVICE EXPERIENCED CLOTTING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: IT WAS REPORTED THAT THE TUBES ARE CLOTTING. WE PROVIDE SUPPLIES AND SERVICES FOR CLINICAL TRIALS RESEARCH. WE WERE NOTIFIED BY ONE OF THE CLINICAL SITES THAT SOME OF CATALOGUE NUMBER 367841 - 2ML K2EDTA BLOOD COLLECTION TUBES WERE CLOTTED. WOULD YOU PLEASE CONFIRM THE EXPECTED STABILITY WINDOW OF WHOLE BLOOD IN THIS BCT AT AMBIENT TEMPERATURE?
IT WAS REPORTED WHEN USING THE BD VACUTAINER® K2 EDTA (K2E) 3.6MG BLOOD COLLECTION TUBES, THE DEVICE EXPERIENCED CLOTTING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: IT WAS REPORTED THAT THE TUBES ARE CLOTTING. WE PROVIDE SUPPLIES AND SERVICES FOR CLINICAL TRIALS RESEARCH. WE WERE NOTIFIED BY ONE OF THE CLINICAL SITES THAT SOME OF CATALOGUE NUMBER 367841 - 2ML K2EDTA BLOOD COLLECTION TUBES WERE CLOTTED. WOULD YOU PLEASE CONFIRM THE EXPECTED STABILITY WINDOW OF WHOLE BLOOD IN THIS BCT AT AMBIENT TEMPERATURE?
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1497849 | BD VACUTAINER® K2 EDTA (K2E) 3.6MG BLOOD COLLECTION TUBES | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON & CO. (BROKEN BOW) | 367841 | SEE H.10. | 50382903678414 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |