FDA Adverse Event Malfunction Summary report: N

X-SMART W/CONTRA ANGLE EUR

MDR report key: 13416356 · Received February 1, 2022

Report

Report Number
8031010-2022-00432
Event Type
Malfunction
Date Received
February 1, 2022
Date of Event
January 17, 2022
Report Date
May 24, 2022
Manufacturer
MAILLEFER INSTRUMENTS HOLDING SARL
Product Code
EGS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THERE HAS BEEN A PREVIOUS REPORT RECEIVED WITH A SIMILAR DEVICE WHERE THIS MALFUNCTION RESULTED IN A SERIOUS INJURY. THEREFORE, IT MUST BE PRESUMED THAT RECURRENCE OF THIS MALFUNCTION COULD POSSIBLY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR REQUIRE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE SUCH. AS SUCH, THIS EVENT IS REPORTABLE PER 21CFR PART 803. THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH HAS NOT BEEN RETURNED AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.

Additional Manufacturer Narrative · 0

MULTIPLE UNSUCCESSFUL ATTEMPTS WERE MADE TO OBTAIN THE DEVICE FOR EVALUATION.

Additional Manufacturer Narrative · 0

DEVICE WAS RECEIVED AND INSPECTED. INSPECTION/INVESTIGATION RESULTS FOR THIS ARE: RECEIVED = 1X X-SMART CARTRIDGE H1004C5210150; 1X X-SMART HEAD CAP H1004C5210500; 1X X-SMART NECK H1004C5470220. X-SMART NECK: OTHER ROTATION NOT FLUID. REPLACED 1X X CARTRIDGE H1004C5210150; 1X X NECK H1004C5470220. 1X X HEAD CAP H1004C5210500

Description of Event or Problem · 0

IN THIS EVENT IT WAS REPORTED THAT AN X-SMART CONTRA ANGLE WON'T HOLD FILES; NO INJURY RESULTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2225261 X-SMART W/CONTRA ANGLE EUR HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL EGS MAILLEFER INSTRUMENTS HOLDING SARL

Patients

Seq Age Sex Outcome Treatment
1 Unknown