HEATER-COOLER SYSTEM 3T
Report
- Report Number
- 9611109-2022-00042
- Event Type
- Malfunction
- Date Received
- February 1, 2022
- Date of Event
- January 4, 2022
- Report Date
- February 1, 2022
- Manufacturer
- LIVANOVA DEUTSCHLAND
- Product Code
- DWC
- UDI-DI
- 04033817901099
- Removal / Correction Number
- Z-2076/2081-2015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THERE WAS NO KNOWN PATIENT INVOLVEMENT. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402). LIVANOVA (B)(4) IMPLEMENTED A FIELD SAFETY NOTICE FOR DISINFECTION AND CLEANING OF HEATER-COOLER DEVICES. THE Z NUMBER IS Z-2076/2081-2015. LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T. THE INCIDENT OCCURRED IN (B)(6). THROUGH FOLLOW-UP COMMUNICATION LIVANOVA LEARNED THAT THE DEVICE WAS CLEANED REGULARLY PER THE INSTRUCTION FOR USE AND THAT IT WAS PLACED INSIDE THE OPERATING THEATER DURING USE WITH THE FAN DIRECTED AWAY FROM THE PATIENT AND AT AN ESTIMATED DISTANCE OF TWO (2) METERS FROM THE SURGERY FIELD. NO ADDITIONAL INFORMATION IS AVAILABLE ON THIS SPECIFIC CASE AND THE ROOT CAUSE COULD NOT BE DETERMINED. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
THERE WAS NO KNOWN PATIENT INVOLVEMENT. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402). LIVANOVA (B)(4) IMPLEMENTED A FIELD SAFETY NOTICE FOR DISINFECTION AND CLEANING OF HEATER-COOLER DEVICES. THE Z NUMBER IS Z-2076/2081-2015. LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T. THE INCIDENT OCCURRED IN (B)(6). THROUGH FOLLOW-UP COMMUNICATION LIVANOVA LEARNED THAT THE DEVICE WAS CLEANED REGULARLY PER THE INSTRUCTION FOR USE AND THAT IT WAS PLACED INSIDE THE OPERATING THEATER DURING USE WITH THE FAN DIRECTED AWAY FROM THE PATIENT AND AT AN ESTIMATED DISTANCE OF TWO (2) METERS FROM THE SURGERY FIELD. NO ADDITIONAL INFORMATION IS AVAILABLE ON THIS SPECIFIC CASE AND THE ROOT CAUSE COULD NOT BE DETERMINED. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
LIVANOVA (B)(4) RECEIVED A REPORT THAT A HEATER-COOLER SYSTEM 3T DEVICE WAS FOUND TO BE CONTAMINATED WITH MYCOBACTERIUM CHIMAERA. THE LABORATORY REPORT CONFIRMING THE REPORTED CONTAMINATION HAS BEEN PROVIDED TO LIVANOVA. THERE IS NO KNOWN PATIENT INVOLVEMENT.
LIVANOVA (B)(4) RECEIVED A REPORT THAT A HEATER-COOLER SYSTEM 3T DEVICE WAS FOUND TO BE CONTAMINATED WITH MYCOBACTERIUM CHIMAERA. THE LABORATORY REPORT CONFIRMING THE REPORTED CONTAMINATION HAS BEEN PROVIDED TO LIVANOVA. THERE IS NO KNOWN PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 993409 | HEATER-COOLER SYSTEM 3T | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | LIVANOVA DEUTSCHLAND | 16-02-80 | 04033817901099 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |