FDA Adverse Event Death Summary report: N

PUMP MMT-1780KPK 670G PATHWAY BLACK MG

MDR report key: 13410833 · Received February 1, 2022

Report

Report Number
2032227-2022-107498
Event Type
Death
Date Received
February 1, 2022
Date of Event
April 7, 2021
Report Date
April 25, 2022
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
000000763000190460
Removal / Correction Number
Z-0956-2020
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

(B)(4). A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. UPDATED ANALYSIS SUMMARY BY (B)(6) ON JAN 27, 2022. S/W 4.11E, RETAINER RING = CLEAR. ON (B)(6) 2021 THE PUMP WAS RETURNED DUE TO PATIENT PASSING AWAY, BUT NOT ALLEGING DEVICE. THE UNIT DID NOT HAVE A BATTERY INSTALLED WHEN RECEIVED. UNIT PASSED THE FUNCTIONAL TEST, INCLUDING THE SELF TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST, DISPLACEMENT TEST AND THE DAT TEST AT 0.0874 INCHES. UNIT UPLOADED PROPERLY USING CARELINK AND THUS. UNIT HAD MINOR SCRATCHED DISPLAY WINDOW, SCRATCHED CASE, CRACKED CASE AT THE CORNER OF THE BELT CLIP RAIL, PILLOWING KEYPAD OVERLAY AND CRACKED RETAINER. THE TEST P-CAP AND RESERVOIR DOES LOCK IN PLACE IN THE RESERVOIR COMPARTMENT. DATA ANALYSIS: THE FORMATTED HISTORY FILE LISTS DATA FROM 01/04/2021 TO 03/25/2021. IN SUMMARY, PUMP PASSED ALL REQUIRED TESTING. PUMP WAS NOT RETURNED FOR AN ALLEGATION.

Additional Manufacturer Narrative · 0

THIS REPORT IS PART OF A RETROSPECTIVE REVIEW AND REMEDIATION EFFORTS IN RESPONSE TO A WARNING LETTER. UPDATED H9: Z-0956-2020. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

IT WAS REPORTED VIA PHONE CALL THAT THE CUSTOMER PASSED AWAY AT HOME ON (B)(6) 2021. THE CAUSE OF DEATH WAS COVID 19. THE CUSTOMER¿S BLOOD GLUCOSE WAS UNKNOWN AT THE TIME OF DEATH. THE CUSTOMER WAS NOT WEARING THE INSULIN PUMP AT THE TIME OF DEATH. NO FURTHER INFORMATION WAS GIVEN. THE INSULIN PUMP WAS RETURNED FOR ANALYSIS. THE CASE IS REPORTED ON BASIS OF FAILURE ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
501997 PUMP MMT-1780KPK 670G PATHWAY BLACK MG ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1780KPK HG3MEMT 000000763000190460

Patients

Seq Age Sex Outcome Treatment
1 60 YR Female Death FRN-MMT-332A-RSVR, UNOMED SET