FDA Adverse Event Injury Summary report: N

HIGH FLOW INSUFFLATION UNIT

MDR report key: 13401816 · Received January 31, 2022

Report

Report Number
8010047-2022-02186
Event Type
Injury
Date Received
January 31, 2022
Date of Event
January 2, 2022
Report Date
May 8, 2024
Manufacturer
SHIRAKAWA OLYMPUS CO., LTD.
Product Code
HIF
UDI-DI
04953170324147
PMA / PMN Number
K122180
Removal / Correction Number
Z-0075-2024
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NO
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE SUSPECT DEVICE HAS BEEN RETURNED TO OLYMPUS FOR EVALUATION AND THE INVESTIGATION IS IN PROCESS. THE OLYMPUS SERVICE CENTER COULD NOT CONFIRM ANY PROBLEM WITH FUNCTIONALITY. THE DEVICE IS WITHOUT ANY SIGN OF DAMAGE. IT IS VERY LIKELY THAT THE DESCRIBED PROBLEM IS CAUSED BY A FAULTY MAIN BOARD WHERE THERE ARE PRESSURE SENSOR CIRCUITS AND CONTROL CIRCUITS PLACED. IT IS A SERIOUS ISSUE AND DUE TO THIS IT IS RECOMMEND TO REPLACE THE MAIN BOARD TO AVOID FURTHER MALFUNCTION. THE DEVICE MEETS ALL LIMITS IN THE INSPECTION PROCEDURE. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED WITH DEVICE EVALUATION RESULTS.

Additional Manufacturer Narrative · 0

THE SUSPECT DEVICE HAS BEEN RETURNED TO OLYMPUS FOR EVALUATION AND THE INVESTIGATION IS IN PROCESS. THE OLYMPUS SERVICE CENTER COULD NOT CONFIRM ANY PROBLEM WITH FUNCTIONALITY. THE DEVICE IS WITHOUT ANY SIGN OF DAMAGE. IT IS VERY LIKELY THAT THE DESCRIBED PROBLEM IS CAUSED BY A FAULTY MAIN BOARD WHERE THERE ARE PRESSURE SENSOR CIRCUITS AND CONTROL CIRCUITS PLACED. IT IS A SERIOUS ISSUE AND DUE TO THIS IT IS RECOMMEND TO REPLACE THE MAIN BOARD TO AVOID FURTHER MALFUNCTION. THE DEVICE MEETS ALL LIMITS IN THE INSPECTION PROCEDURE. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED WITH DEVICE EVALUATION RESULTS.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. IT HAS BEEN OVER 7 YEARS SINCE THE SUBJECT DEVICE WAS MANUFACTURED. BASED ON THE RESULTS OF THE INVESTIGATION, HE CAUSE OF IT COULD NOT BE IDENTIFIED BECAUSE NO ABNORMALITIES WERE FOUND IN THE DEVICE INSPECTION. OLYMPUS WILL CONTINUE TO MONITOR THE FIELD PERFORMANCE OF THIS DEVICE.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUPPLEMENTED TO CORRECT H6 PROBLEM CODE. ADDITIONAL INFORMATION PROVIDED IN B2, H6 HEALTH CLINICAL CODE, H6 HEALTH IMPACT CODE, H7, H9. A FORMAL INVESTIGATION WAS INITIATED TO INVESTIGATE THE ROOT CAUSE. BASED ON THE RESULTS OF THE INVESTIGATION, A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED AT THIS TIME. HOWEVER, THE FOLLOWING ARE CONSIDERED TO BE POTENTIALLY RELATED TO THE ROOT CAUSE OF THE EVENT: -MODERATE OVERPRESSURE OCCURRED DUE TO DEVICE FAILURE. -OVERPRESSURE DUE TO THE SMALL ABDOMINAL CAVITY. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED TO CORRECT THE LEGAL MANUFACTURER¿S CONTACT INFORMATION AND FACILITY REGISTRATION NUMBER. THE FACILITY REGISTRATION NUMBER IS 3002808148.

Description of Event or Problem · 0

THE CUSTOMER REPORTED TO OLYMPUS, DURING A THERAPEUTIC LAPAROSCOPIC LIGATURE OF COMMUNICATING HYDROCELE, THE SUBJECT DEVICE MALFUNCTIONED. THE PRESSURE ON THE INSUFFLATOR WAS SET TO 8 BUT THE MEASURED PRESSURE WAS AT 24 AND THE PROCEDURE WAS ABORTED. EARLY IN THE PROCEDURE, THE ANESTHESIA NURSE REPORTED INCREASED EFFORTS WERE NEEDED TO VENTILATED THE PATIENT ADEQUATELY. THE ABDOMINAL PRESSURE WAS LOWERED FROM 12-10MMHG. THE PATIENT QUICKLY DESATURATED TO 60%. THE ANESTHESIA MONITORING EQUIPMENT WAS ALARMING 'HIGH.' AT THAT TIME IS WHEN IT WAS IDENTIFIED THE INSUFFLATOR HAD BLOWN THE ABDOMINAL PRESSURE TO 24MMHG DESPITE THE SET PRESSURE LIMIT OF 10MMHG. THE GAS INSUFFLATOR ALSO ALARMED BUT IT WAS DIFFICULT TO HEAR GIVEN THE HIGH ALARMS FROM THE ANESTHESIA EQUIPMENT. THE GAS HOSE WAS IMMEDIATELY DISCONNECTED FROM THE PATIENT AND THE PATIENT'S STATUS NORMALIZED QUICKLY. A REVIEW OF THE LOG INDICATED THE DURATION OF SEVERE DESATURATION WAS 20 SECONDS. THE SUBJECT DEVICE WAS RESTARTED AND ALL THE SETTINGS WERE VERIFIED TO BE CORRECT INCLUDING SMALL CAVITY SELECTION. THE FACILITY PROCEEDED TO TEST THE DEVICE WITH LOW FLOW AND A PRESSURE LIMIT OF 8MMHG. THE ABDOMINAL PRESSURE INCREASED AGAIN WELL ABOVE THE LIMIT AND THE DEVICE WAS DISCONNECTED WHEN THE PRESSURE REACHED 15MMHG. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE AND THERE WAS NO FURTHER ISSUES. THE PROCEDURAL DELAY WAS NOT REPORTED. IT WAS ALSO REPORTED THE PATIENT HAD AN HIGH BLOOD PRESSURE EVENT DURING THE PROCEDURE AS WELL. NO DETAILS WERE PROVIDED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED TO OLYMPUS, DURING A THERAPEUTIC LAPAROSCOPIC LIGATURE OF COMMUNICATING HYDROCELE, THE SUBJECT DEVICE MALFUNCTIONED. THE PRESSURE ON THE INSUFFLATOR WAS SET TO 8 BUT THE MEASURED PRESSURE WAS AT 24 AND THE PROCEDURE WAS ABORTED. EARLY IN THE PROCEDURE, THE ANESTHESIA NURSE REPORTED INCREASED EFFORTS WERE NEEDED TO VENTILATED THE PATIENT ADEQUATELY. THE ABDOMINAL PRESSURE WAS LOWERED FROM 12-10MMHG. THE PATIENT QUICKLY DESATURATED TO 60%. THE ANESTHESIA MONITORING EQUIPMENT WAS ALARMING 'HIGH.' AT THAT TIME IS WHEN IT WAS IDENTIFIED THE INSUFFLATOR HAD BLOWN THE ABDOMINAL PRESSURE TO 24MMHG DESPITE THE SET PRESSURE LIMIT OF 10MMHG. THE GAS INSUFFLATOR ALSO ALARMED BUT IT WAS DIFFICULT TO HEAR GIVEN THE HIGH ALARMS FROM THE ANESTHESIA EQUIPMENT. THE GAS HOSE WAS IMMEDIATELY DISCONNECTED FROM THE PATIENT AND THE PATIENT'S STATUS NORMALIZED QUICKLY. A REVIEW OF THE LOG INDICATED THE DURATION OF SEVERE DESATURATION WAS 20 SECONDS. THE SUBJECT DEVICE WAS RESTARTED AND ALL THE SETTINGS WERE VERIFIED TO BE CORRECT INCLUDING SMALL CAVITY SELECTION. THE FACILITY PROCEEDED TO TEST THE DEVICE WITH LOW FLOW AND A PRESSURE LIMIT OF 8MMHG. THE ABDOMINAL PRESSURE INCREASED AGAIN WELL ABOVE THE LIMIT AND THE DEVICE WAS DISCONNECTED WHEN THE PRESSURE REACHED 15MMHG. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE AND THERE WAS NO FURTHER ISSUES. THE PROCEDURAL DELAY WAS NOT REPORTED. IT WAS ALSO REPORTED THE PATIENT HAD AN HIGH BLOOD PRESSURE EVENT DURING THE PROCEDURE AS WELL. NO DETAILS WERE PROVIDED.

Description of Event or Problem · 0

THE CUSTOMER CONFIRMED THERE WERE NO CHANGE OF TROCAR POSITION NOR ANY OTHER CHANGES IN ADDITION TO UHI EXCHANGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1641648 HIGH FLOW INSUFFLATION UNIT HIGH FLOW INSUFFLATION UNIT HIF SHIRAKAWA OLYMPUS CO., LTD. UHI-4 04953170324147

Patients

Seq Age Sex Outcome Treatment
1 3 YR Male Other| R