ABL80 FLEX CO-OX ANALYZER

MDR report key: 1340168 · Received March 5, 2009

Report

Report Number
2027541-2009-00002
Event Type
Malfunction
Date Received
March 5, 2009
Date of Event
January 1, 2009
Report Date
February 6, 2009
Manufacturer
SENDX MEDICAL, INC.
Product Code
CHL
PMA / PMN Number
K080370
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US

Narratives

Additional Manufacturer Narrative · 1

THE ALLEGED FAILURE COULD NOT BE DUPLICATED. THE DEVICE IS DESIGNED TO PREVENT ISSUES SIMILAR TO THE ALLEGED FAILURE. NUMEROUS ATTEMPTS HAVE BEEN MADE REQUESTING THAT THE HOSPITAL DESCRIBE THE INCIDENT IN GREATER DETAIL. NONE OF THE RESPONSES FROM THE HOSPITAL HAVE PROVIDED SUFFICIENT INFO TO REPLICATE THE EXACT CIRCUMSTANCES THAT LEAD TO THE ALLEGED FAILURE. ONE RESPONSE INDICATED THAT AT THE TIME OF THE EVENT THE ANALYZER WAS NOT BEING USED AND WAS LOCATED IN A CONFERENCE ROOM. QUALITY CONTROL SOLUTION ALLEGEDLY SPRAYED THE HOSPITAL EMPLOYEE. THE FLUID PATH TAKEN BY A QUALITY CONTROL SOLUTION IS NEARLY IDENTICAL TO THE FLUID PATH TAKEN BY A BLOOD SAMPLE. IF THE SAME ALLEGED MALFUNCTION OCCURRED WITH BLOOD INSTEAD OF NON-HAZARDOUS QUALITY CONTROL SOLUTION, BLOOD COULD HAVE SPRAYED THE HOSPITAL EMPLOYEE. EVALUATION SUMMARY: THE ABL80 FLEX CO-OX ANALYZER (B) (4) WAS ANALYZED BY THE DISTRIBUTOR, (B) (4). IN AN ATTEMPT TO DUPLICATE THE SPRAYING INCIDENT: THE WASTE MANIFOLD WAS COMPLETELY BLOCKED, RESULTING IN THE WASTE PUMP TUBING DISCONNECTING AND LEAKING UNDERNEATH THE SOLUTION PACK. THE WASTE PUMP TUBING WAS TIED OFF WITH A ZIP TIE. THIS FORCED THE SOLUTION TO BACK UP AND LEAK OUT OF THE SENSOR CASSETTE WASTE PORT. IN BOTH OF THE ABOVE CASES, THE OUTCOME WAS A LEAK OF SOLUTION FROM THE ANALYZER. SOLUTION DID NOT SPRAY FROM THE ANALYZER. THE ANALYZER FUNCTIONED AS INTENDED.

Description of Event or Problem · 1

ON 02/06/2009, A DISTRIBUTOR REPORTED THAT A COMPLAINT WAS RECEIVED INDICATING THE UNIT HAD A LEAK THAT SPRAYED A HOSPITAL EMPLOYEE DURING A VIDEO CONFERENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ABL80 FLEX CO-OX ANALYZER PH, BLOODGAS, ELECTROLYTE ANALYSIS SYSTEM CHL SENDX MEDICAL, INC. 393-841