MICROSTAAR INJECTOR
Report
- Report Number
- 2023826-2009-00203
- Event Type
- Malfunction
- Date Received
- March 6, 2009
- Date of Event
- January 15, 2009
- Report Date
- February 6, 2009
- Manufacturer
- STAAR SURGICAL CO.
- Product Code
- KYB
- PMA / PMN Number
- K954600
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SF
- Reporter Occupation
- PHYSICIAN
Narratives
CONCLUSION: (OTHER) - AN INVESTIGATION WAS OPENED TO EVALUATE A COMPLAINT TREND ASSOCIATED WITH LENS TEARS THAT WAS ORIGINALLY IDENTIFIED IN JUNE 2005. POSSIBLE ROOT CAUSE FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND POSSIBLE HANDLING ERRORS BY THE CUSTOMER. TO ADDRESS DELIVERY SYSTEM ISSUES, ALL STAGES IN THE MANUFACTURING OF THE INJECTORS AND CARTRIDGES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. TO ADDRESS HANDLING ERRORS, ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.
IT WAS REPORTED THAT THE SURGEON INSERTED AN AA4204VL SILICONE SINGLE PIECE LENS AND THE PLUNGER OF THE INJECTOR CAUGHT AND TORE THE TRAILING HAPTIC UPON INSERTION. THE INCISION WAS ENLARGED TO REMOVE THE LENS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MICROSTAAR INJECTOR | INTRAOCULAR LENS GUIDE | KYB | STAAR SURGICAL CO. | MSI-PM | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Required Intervention | LENS: MODEL AA4204VL| CARTRIDGE: MODEL AQ CARTRIDGE-FP |