FDA Adverse Event Malfunction Summary report: N

MICROSTAAR INJECTOR

MDR report key: 1338662 · Received March 6, 2009

Report

Report Number
2023826-2009-00203
Event Type
Malfunction
Date Received
March 6, 2009
Date of Event
January 15, 2009
Report Date
February 6, 2009
Manufacturer
STAAR SURGICAL CO.
Product Code
KYB
PMA / PMN Number
K954600
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SF
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

CONCLUSION: (OTHER) - AN INVESTIGATION WAS OPENED TO EVALUATE A COMPLAINT TREND ASSOCIATED WITH LENS TEARS THAT WAS ORIGINALLY IDENTIFIED IN JUNE 2005. POSSIBLE ROOT CAUSE FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND POSSIBLE HANDLING ERRORS BY THE CUSTOMER. TO ADDRESS DELIVERY SYSTEM ISSUES, ALL STAGES IN THE MANUFACTURING OF THE INJECTORS AND CARTRIDGES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. TO ADDRESS HANDLING ERRORS, ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SURGEON INSERTED AN AA4204VL SILICONE SINGLE PIECE LENS AND THE PLUNGER OF THE INJECTOR CAUGHT AND TORE THE TRAILING HAPTIC UPON INSERTION. THE INCISION WAS ENLARGED TO REMOVE THE LENS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICROSTAAR INJECTOR INTRAOCULAR LENS GUIDE KYB STAAR SURGICAL CO. MSI-PM UNK

Patients

Seq Age Sex Outcome Treatment
1 59 YR Required Intervention LENS: MODEL AA4204VL| CARTRIDGE: MODEL AQ CARTRIDGE-FP