FDA Adverse Event Injury Summary report: N

PHILIPS MEDICAL SYSTEMS

MDR report key: 1338207 · Received March 6, 2009

Report

Report Number
MW5010223
Event Type
Injury
Date Received
March 6, 2009
Date of Event
February 6, 2009
Report Date
March 4, 2009
Product Code
DRG
Adverse Event
Yes
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
CO, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

A PHILIPS TELEMETRY TRANSMITTER BECAME VERY HOT WHILE IN USE ON A PATIENT. THE NURSING STAFF NOTICED THE SMELL OF THE OVERHEATED TRANSMITTER. NURSING STAFF REPLACED THE TRANSMITTER AND THE BIOMED WAS ABLE TO EXAMINE IT ON THE NEXT BUSINESS DAY. SOME SMALL PLASTIC FEATURES INSIDE THE BATTERY COMPARTMENT WERE FOUND TO BE BROKEN OFF AND THIS ALLOWED THE SPRING BATTERY CONTACTS TO BE DISLOCATED IN A WAY THAT ALLOWED THE NEGATIVE CONTACT TO TOUCH THE POSITIVE CONTACT. THIS SHORT CIRCUITED THE BATTERY, OVERHEATING THE BATTERY AND THE TRANSMITTER. THE DESIGN OF THE BATTERY ENCLOSURE AND SPRING CONTACTS MAKES THESE DEVICES EASILY SUSCEPTIBLE TO THIS FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PHILIPS MEDICAL SYSTEMS TELEMETRY TRANSMITTER DRG M2601B

Patients

Seq Age Sex Outcome Treatment
1