INFANT INTENSIVE CARE SYSTEM
MDR report key: 133769
·
Received November 21, 1997
Report
- Report Number
- 2510954-1997-00013
- Event Type
- Other
- Date Received
- November 21, 1997
- Date of Event
- October 21, 1997
- Report Date
- October 22, 1997
- Manufacturer
- AIR- SHIELDS
- Product Code
- FMT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
Description of Event or Problem · 1
IT WAS REPORTED THAT A NURSE NOTICED A SPARK AT THE TOP OF THE RADIANT WARMER AND WITNESSED A LITTLE "LIGHTNING BOLT" (ARC) ENTERING THE PTS CHEST AND EXITING THROUGH THE RIGHT HAND. THEN THE MATTRESS COVERING CAUGHT FIRE AND THE INFANT WAS REMOVED. THE INFANT RECEIVED MINOR BURNS ON THE CHEST AND FOUR FINGERS. NO CARDIAC ARREST OR OTHER PROBLEMS WITH THE PT WERE NOTED AND THE INFANT APPEARED TO BE IN GOOD CONDITION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFANT INTENSIVE CARE SYSTEM | INFANT RADIANT WARMER | FMT | AIR- SHIELDS | 7810 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |