G7 BALL HEX DRVR FOR INSR HNDL
Report
- Report Number
- 0001825034-2022-00195
- Event Type
- Malfunction
- Date Received
- January 27, 2022
- Date of Event
- January 4, 2022
- Report Date
- March 22, 2022
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- LPH
- UDI-DI
- 00880304926820
- PMA / PMN Number
- K121874
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
(B)(4). PRODUCT HAS BEEN RECEIVED BY ZIMMER BIOMET AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: ITEM # 110010247/ G7 OSSEOTI 4 HOLE SHELL 58MM/ LOT # UNKNOWN. ITEM # 110003451/ G7 STR MODULAR SHELL INSERTER/ LOT # 494810. ITEM # 110003452/ G7 STR INSRTR THREADED SHAFT/ LOT # 072082. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034 -2022 -00197, 0001825034 -2022 -00199, AND 0001825034 -2022 -00200.
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. VISUAL INSPECTION OF THE RETURNED HANDLE FOUND THE BALL HEX TRIP TO BE FRACTURED. OXIDATION SPOTS WERE OBSERVED ON THE SHAFT AND GRIP. SURFACE SCRATCHES ARE ALSO PRESENT ON THE SHAFT. DHR WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. THE ROOT CAUSE IS UNABLE TO BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT WAS REPORTED THAT DURING A HIP PROCEDURE THE CUP INSERTION HANDLE COULD NOT BE LOOSENED FROM CUP. TOMMY BAR WAS USED WITH G7 BALL TIP SCREW DRIVER TO TRY TO RELEASE CUP. G7 BALL DRIVER BROKE AT TIP AND INSERTION HANDLE WAS STILL ATTACHED TO CUP IN PATIENT. WHOLE CONSTRUCT WAS REMOVED. NO ADVERSE EVENTS WERE REPORTED. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 112680 | G7 BALL HEX DRVR FOR INSR HNDL | INSTRUMENT, HIP | LPH | ZIMMER BIOMET, INC. | NI | 062100 | 00880304926820 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR | Male |