FDA Adverse Event Malfunction Summary report: N

PUMP MMT-1781K 670G V4.11 MG

MDR report key: 13364360 · Received January 27, 2022

Report

Report Number
2032227-2022-106263
Event Type
Malfunction
Date Received
January 27, 2022
Date of Event
January 23, 2022
Report Date
July 12, 2022
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
000000763000283544
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 0

RETAINER RING=BLACK. THE PUMP WAS RETURNED FOR PRIME/FILL ANOMALY FOUND ON (B)(6) 2022. PUMP¿S HISTORY AND TRACE FILES DOWNLOADED SUCCESSFULLY USING THUS SOFTWARE. UNIT PASSED DISPLACEMENT TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST, OCCLUSION TEST, AND SELF-TEST. NO PRIME FILL ANOMALY NOTED DURING TESTING. TESTING. NO PRIME/FILL ALARMS NOTED IN THE PUMP HISTORY/TRACE FILES. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND NO EVIDENCE OF PHYSICAL OR MOISTURE DAMAGE ON THE ELECTRONIC ASSEMBLY, MOTOR OR FORCE SENSOR. FORCE SENSOR WAS MEASURED AND IS WITHIN SPEC 23.8 MV AT ZERO OFFSET. NO DAMAGE NOTED AT THE ELECTRONIC ASSEMBLIES DURING VISUAL INSPECTION. TEST P-CAP / RESERVOIR LOCKS PROPERLY INTO PLACE. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: SCRATCHED CASE AND CRACKED CASE ON THE BATTERY TUBE SIDE. PRIME/FILL ANOMALY NOT CONFIRMED DURING TESTING. THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER WAS UNABLE TO EXIT LOAD RESERVOIR PROCESS. TROUBLE SHOOTING WAS PERFORMED. CUSTOMER DID NOT RECEIVE COMPLETE STATUS WITH NEXT ON THE SCREEN. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
40887 PUMP MMT-1781K 670G V4.11 MG ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1781K HG46TDK 000000763000283544

Patients

Seq Age Sex Outcome Treatment
1 Unknown