FDA Adverse Event Injury Summary report: N

CONSERVE® A-CLASS® BFH® HEAD MEDIUM NECK 44MM

MDR report key: 13357702 · Received January 26, 2022

Report

Report Number
3010536692-2022-00022
Event Type
Injury
Date Received
January 26, 2022
Report Date
January 26, 2022
Manufacturer
MICROPORT ORTHOPEDICS INC.
Product Code
KWA
UDI-DI
M68438AM44001
PMA / PMN Number
K021349
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Description of Event or Problem · 0

ALLEGEDLY, PATIENT STARTED TO EXPERIENCE PAIN IN OR AROUND HIS RIGHT HIP JOINT, ALONG WITH COBALT AND CHROMIUM LEVELS THAT WERE INCREASING OVER THE PAST FEW YEARS. CONTINUING AND INCREASING PAIN IN HIS RIGHT HIP, AND ELEVATED SERUM COBALT/CHROMIUM LEVELS REPORTED ON JULY 22, 2019. IN (B)(6) 2019, RIGHT HIP REVISION PROCEDURE PATIENT´S RIGHT HIP WAS REVISED, RESULTING IN REMOVAL AND REPLACEMENT OF THE PREVIOUSLY IMPLANTED CONSERVE-DYNASTY-PROFEMUR HIP SYSTEM METAL-ON-METAL BEARING SURFACE WITH A BIOMET DUAL MOBILITY ACTIVE ARTICULATION HIP SYSTEM, AND A MICROPORT ALUMINA MATRIC COMPOSITE DELTA OPTION HEAD, AND A DELTA OPTION NECK SLEEVE. AN OPERATIVE NOTE FROM (B)(6) 2019, REVISION PROCEDURE REFERENCES A FINDING OF STAINED SYNOVIUM AND SIGNIFICANT CORROSION ON THE MODULAR NECK. THE EXTRACTED MODULAR NECK SHOWS SIGNIFICANT CORROSION ON BOTH THE PROXIMAL AND DISTAL TRUNNIONS OF THAT DEVICE. DUE TO CONCERNS WITH A HISTORY OF THE WRIGHT MEDICAL TITANIUM PROFEMUR MODULAR NECKS CORRODING, AND FRACTURING, THE INTRA-OPERATIVE MEDICAL DECISION WAS ALSO MADE BY SURGEON TO REMOVE AND REPLACE THE TITANIUM PROFEMUR MODULAR NECK OF THAT HIP WITH A NEW, NOT YET CORRODING, TITANIUM MODULAR NECK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
39724 CONSERVE® A-CLASS® BFH® HEAD MEDIUM NECK 44MM HIP COMPONENT KWA MICROPORT ORTHOPEDICS INC. 38AM4400 097471657 M68438AM44001

Patients

Seq Age Sex Outcome Treatment
1 Male Required Intervention