ADVIA CENTAUR MYOGLOBIN (MYO)
Report
- Report Number
- 1219913-2009-00022
- Event Type
- Other
- Date Received
- March 4, 2009
- Date of Event
- February 17, 2009
- Report Date
- February 17, 2009
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- DDR
- PMA / PMN Number
- K974325
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A SIEMENS FIELD SERVICE ENGINEER (FSE) WAS SENT TO THE CUSTOMER SITE. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATED THAT THE SAMPLE PROBE VERTICAL PULLEY NEEDED ADJUSTMENT. THE FSE READJUSTED AND TIGHTENED THE VERTICAL PULLEY AND RECALIBRATED THE SAMPLE PROBE. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVAL OF THE DEVICE IS REQUIRED.
TWO FALSELY SUPPRESSED ADVIA CENTAUR MYOGLOBIN RESULTS WERE OBTAINED ON PT SAMPLES. REPEAT TESTING WAS PERFORMED ON THE PT SAMPLES AND THE TEST RESULTS WERE HIGHER. THE SECOND PT SAMPLE WAS TESTED ON THE ADVIA CENTAUR CP AND THE RESULT WAS MUCH HIGHER THAN THE REPEAT ON THE ADVIA CENTAUR. PT TREATMENT WAS NOT PRESCRIBED OR ALTERED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT MYOGLOBIN RESULTS.
A SIEMENS FIELD SERVICE ENGINEER (FSE) WAS SENT TO THE CUSTOMER SITE. ANALYSIS OF THE INSTRUMENT AND VERTICAL PULLEY NEEDED ADJUSTMENT. THE FSE READJUSTED AND TIGHTENED THE VERTICAL PULLEY AND RECALIBRATED THE SAMPLE PROBE. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVAL OF THE DEVICE IS REQUIRED.
TWO FALSELY SUPPRESSED ADVIA CENTAUR MYOGLOBIN RESULTS WERE OBTAINED ON PT SAMPLES. REPEAT TESTING WAS PERFORMED ON THE PT SAMPLES AND THE TEST RESULTS WERE HIGHER. THE SECOND PT SAMPLE WAS TESTED ON THE ADVIA CENTAUR CP AND THE RESULT WAS MUCH HIGHER THAN THE REPEAT ON THE ADVIA CENTAUR. PT TREATMENT WAS NOT PRESCRIBED OR ALTERED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT MYOGLOBIN RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADVIA CENTAUR MYOGLOBIN (MYO) | MYOGLOBIN IMMUNOASSAY | DDR | SIEMENS HEALTHCARE DIAGNOSTICS INC. | NA | 055160 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ||||
| 2 |