6308 2000ML DRAIN BG W SPOUT
Report
- Report Number
- 9612030-2022-03143
- Event Type
- Malfunction
- Date Received
- January 25, 2022
- Date of Event
- January 24, 2022
- Report Date
- April 19, 2022
- Manufacturer
- CARDINAL HEALTH
- Product Code
- KNX
- UDI-DI
- 10884521002050
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
AN INVESTIGATION IS CURRENTLY UNDERWAY. UPON COMPLETION, THE RESULTS WILL BE FORWARDED.
SINCE A LOT NUMBER WAS NOT PROVIDED, THE DEVICE HISTORY RECORD COULD NOT BE PERFORMED. AS PART OF OUR MANUFACTURING PROCESS, ALL DEVICE HISTORY RECORDS ARE REVIEWED AND APPROVED BY QUALITY, PRIOR TO RELEASE OF PRODUCT. ONE USED SAMPLE WITHOUT ITS ORIGINAL PACKAGING WAS RECEIVED FOR THE EVALUATION. A VISUAL INSPECTION OF THE RECEIVED SAMPLE WAS PERFORMED, IN ADDITIONAL A FUNCTIONAL TEST WAS CARRIED OUT. THE BAG WAS FILLED TO ITS MAXIMUM CAPACITY (2000 ML). THE INVESTIGATION WAS CARRIED OUT WITH THE MULTIFUNCTIONAL TEAM. UPON EVALUATION, THE REPORTED PROBLEM COULD NOT BE CONFIRMED. THE BAG WAS FILLED TO ITS MAXIMUM CAPACITY WITHOUT OBSERVING ANY CONDITION THAT PREVENTED ITS FILLING. ALL PROCESSES WERE REVIEWED. ALL PROCESSES AND CONTROLS WERE FOUND TO BE PROPERLY FOLLOWED, INCLUDING PACKAGING AND ALL INSPECTIONS PERFORMED ON THE PRODUCT. NO ABNORMAL CONDITIONS WERE FOUND THAT COULD TRIGGER THE REPORTED CONDITION. WE WILL KEEP MONITORING THE PROCESS FOR ANY ADVERSE TRENDS THAT REQUIRE IMMEDIATE ATTENTION. THIS COMPLAINT WILL BE USED FOR QA TRACKING AND TRENDING PURPOSES.
THE CUSTOMER REPORTED THAT CLOGGING OCCURRED AT THE ROUND PART WHERE THE TUBE AND BAG ARE CONNECTED. THERE WAS NO PATIENT HARM. THE DURATION OF USE IS UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 831374 | 6308 2000ML DRAIN BG W SPOUT | COLLECTOR, URINE, (AND ACCESSORIES) FOR INDWELLING CATHETER | KNX | CARDINAL HEALTH | 6308 | 10884521002050 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |