FDA Adverse Event Malfunction Summary report: N

T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY

MDR report key: 13343097 · Received January 25, 2022

Report

Report Number
3013756811-2022-04608
Event Type
Malfunction
Date Received
January 25, 2022
Date of Event
January 1, 2022
Report Date
January 25, 2022
Manufacturer
TANDEM DIABETES CARE
Product Code
QFG
UDI-DI
00850006613786
PMA / PMN Number
K201214
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT INSULIN DRIPS WERE NOT OBSERVED TO BE EXITING THE INFUSION SET TUBING DURING THE LOAD FILL TUBING PROCESS. REPORTEDLY, A SUPPLY CHANGE WAS PERFORMED TO ADDRESS THE ISSUE AND INSULIN DELIVERY WAS RESUMED. CUSTOMER'S BLOOD GLUCOSE WAS 160-208 MG/DL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1094338 T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY ALTERNATE CONTROLLER ENABLED INFUSION PUMP QFG TANDEM DIABETES CARE 1002717 60262316 00850006613786

Patients

Seq Age Sex Outcome Treatment
1 64 YR Male INSULIN: NOVOLOGINFUSION SET: AUTOSOFT XC