FDA Adverse Event
Malfunction
Summary report: N
DISCOVERY MR 750 3.0T
MDR report key: 13341780
·
Received January 25, 2022
Report
- Report Number
- 13341780
- Event Type
- Malfunction
- Date Received
- January 25, 2022
- Date of Event
- January 12, 2022
- Report Date
- January 13, 2022
- Manufacturer
- GENERAL ELECTRIC COMPANY
- Product Code
- LNH
- UDI-DI
- 00840682115872
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- MN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
Description of Event or Problem · 0
MR'S CONTROL PANEL HAD AN ELECTRICAL SHORTAGE THOUGHT TO BE FROM THE USE OF SATURATED CLEANING WIPES WHICH CAUSED CORROSION. THIS MORNING WHEN ENTERING THE SCAN ROOM TO RUN THE MORNING ACR, TECHS NOTICED THE ROOM HAD AN ODD ODOR. IT WAS THEN NOTICED THAT THE CONTROL PANEL HAD MELTING PLASTIC AND EXTREME HEAT COMING FROM IT. SERVICE WAS NOTIFIED AND THE EMERGENCY POWER OFF WAS PRESSED. SERVICE CAME, EVALUATED, AND FIXED THE SCANNER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 662170 | DISCOVERY MR 750 3.0T | SYSTEM, NUCLEAR MAGNETIC RESONANCE IMAGING | LNH | GENERAL ELECTRIC COMPANY | DISCOVERY MR750 3.0T | 00840682115872 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |