FDA Adverse Event Malfunction Summary report: N

MINIMED 770G US SYSTEM BLE CONNECT 3.0 MG/DL

MDR report key: 13341190 · Received January 25, 2022

Report

Report Number
2032227-2022-105906
Event Type
Malfunction
Date Received
January 25, 2022
Date of Event
January 21, 2022
Report Date
June 13, 2022
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
000000763000414344
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 0

RETAINER RING = BLACK CUSTOMER RETURNED INSULIN PUMP FOR AN ALLEGED CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM FOUND ON (B)(6) 2022. INSULIN PUMP WAS RECEIVED WITH A CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. UNABLE TO PERFORM THE DISPLACEMENT TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST, OCCLUSION TEST, SLEEP CURRENT MEASUREMENT TEST, ACTIVE CURRENT MEASUREMENT TEST AND SELF TEST DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. SUCCESSFULLY DOWNLOADED TRACES AND HISTORY FILE USING THUMP. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM DUE TO PUMP ERROR 53 ALARM AND PUMP ERROR 3 ALARM FOUND ON (B)(6) 2022 02:40:00.000 AND (B)(6) 2022 02:47:10.000. AS PER GLOBAL LOGIC ANALYSIS (ESF#:3923533), PUMP ERROR 53 ALARMS (LINE NUMBER 265 FILE NUMBER 32) AND PUMP ERROR 3 ALARMS (LINE NUMBER 1541 FILE NUMBER 2002), SUSPECTED ON HW. INSULIN PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND NO EVIDENCE OF PHYSICAL OR MOISTURE DAMAGE ON THE ELECTRICAL BOARD 1, ELECTRICAL BOARD 2, FORCE SENSOR, MOTOR AND VIBRATOR ASSEMBLY NOTED. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE ALSO NOTED DURING VISUAL INSPECTION: A PILLOWING KEYPAD OVERLAY AND A SCRATCHED CASE. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM CONFIRMED DUE TO PUMP ERROR 53 ALARM AND PUMP ERROR 3 ALARM. PER R&D ENGINEER, CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM, PUMP ERROR 53 ALARM AND PUMP ERROR 3 ALARM, SUSPECTED ON HARDWARE. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A CRITICAL PUMP ERROR ALARM AND SHOW RED X ON THE SCREEN. INSULIN PUMP HAD OPEN BOOK IMAGE DISPLAYED ON THE SCREEN. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1596693 MINIMED 770G US SYSTEM BLE CONNECT 3.0 MG/DL ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1880 HG4SH4GZZ 000000763000414344

Patients

Seq Age Sex Outcome Treatment
1 27 YR Male