FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 13340174 · Received January 25, 2022

Report

Report Number
9610877-2022-50155
Event Type
Malfunction
Date Received
January 25, 2022
Date of Event
January 11, 2022
Report Date
January 25, 2022
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
EOB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL VNL11-J10 IS AVAILABLE IN THE USA WITH A 510K NUMBER K172156. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD MODULE FLUID DAMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE FLUID DAMAGE FROM THE CCD MODULE. IN ADDITION, WE CONFIRMED THAT THE INSERTION FLEXIBLE TUBE (IFT) DEFORMED AND LEAKY; HOWEVER, THEY ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (FLUID DAMAGE).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1224971 PENTAX VIDEO NASO PHARYNGO LARYNGOSTROBOSCOPE EOB HOYA CORPORATION PENTAX TOKYO OFFICE VNL-1190STK

Patients

Seq Age Sex Outcome Treatment
1 Unknown