FDA Adverse Event Malfunction Summary report: N

VMP ENDURANCE 80G

MDR report key: 13339375 · Received January 24, 2022

Report

Report Number
1818910-2022-01530
Event Type
Malfunction
Date Received
January 24, 2022
Date of Event
January 11, 2022
Report Date
January 24, 2022
Manufacturer
DEPUY ORTHOPAEDICS INC US
Product Code
HTQ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY : NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY. DEVICE HISTORY LOT : A MANUFACTURING RECORD EVALUATION (NC SEARCH) WAS PERFORMED FOR THE FINISHED DEVICE 3172080/9783780 NUMBER, AND NO NON-CONFORMANCES / MANUFACTURING IRREGULARITIES RELATED TO THE MALFUNCTION WERE IDENTIFIED.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY : THE DEVICE ASSOCIATED WITH THIS REPORT WAS RETURNED FOR ANALYSIS. VISUAL EXAMINATION OF THE PART EXHIBITED FILTER CONNECTOR BROKEN AND SEPARATED FROM THE REST OF THE FILTER. THE DEVICE'S PACKAGE WAS NOT RETURNED FOR EXAMINATION. ADDITIONALLY, THERE WERE OBSERVED FLUIDS TRACES THROUGH THE HOSE'S INTERNAL LENGTH WHICH IS INDICATIVE OF HAVE BEEN MANIPULATED DURING PROCEDURE. DEPUY SYNTHES CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION MAY BE RE-OPENED AS NECESSARY. DEVICE HISTORY LOT : A MANUFACTURING RECORD EVALUATION (NC SEARCH) WAS PERFORMED FOR THE FINISHED DEVICE 3172080/9783780, AND NO NON-CONFORMANCES/MANUFACTURING IRREGULARITIES RELATED TO THE MALFUNCTION WERE IDENTIFIED.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H10 ADDITIONAL NARRATIVE: ADDED: D9, H4. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. CORRECTED: H3.

Description of Event or Problem · 0

IT WAS REPORTED THAT ON (B)(6) 2022, THE PATIENT UNDERWENT THE SURGERY WITH THE VACU-MIX PLUS CEMENT IN QUESTION. IN THE SURGERY, WHEN THE SURGEON INSERTED THE TUBE OF THE MIXING SYSTEM INTO THE VACUUM PUMP, THE TIP WAS CRACKED. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH A NEW PRODUCT. NO FURTHER INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1093533 VMP ENDURANCE 80G BONE CEMENT : BONE CEMENT HTQ DEPUY ORTHOPAEDICS INC US 9783780

Patients

Seq Age Sex Outcome Treatment
1 Unknown