BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM
Report
- Report Number
- 3014704491-2022-00031
- Event Type
- Malfunction
- Date Received
- January 24, 2022
- Date of Event
- November 15, 2021
- Report Date
- February 6, 2022
- Manufacturer
- BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
- Product Code
- FOZ
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
H6: INVESTIGATION SUMMARY A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 1050039. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, VISUAL ANALYSIS WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RETAINED DEVICES WERE FOUND TO BE FREE OF ANY DAMAGE OR OTHER ABNORMALITIES. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT. H3 OTHER TEXT : SEE H10.
IT WAS REPORTED THAT BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM HAD THE CATHETER KINK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "IN THE PROCESS OF VENIPUNCTURE, THE DISCONTINUOUS PART OF THE INDWELLING NEEDLE WAS FOUND TO BE TWISTED, SO IT WAS REPLACED."
IT WAS REPORTED THAT BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM HAD THE CATHETER KINK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "IN THE PROCESS OF VENIPUNCTURE, THE DISCONTINUOUS PART OF THE INDWELLING NEEDLE WAS FOUND TO BE TWISTED, SO IT WAS REPLACED"
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 687741 | BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM | INTRAVASCULAR CATHETER | FOZ | BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU | 1050039 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |