BD PERFUSION 50 ML SYRINGE WITH MOUNTED NEEDLE, OPAQUE
Report
- Report Number
- 3003916417-2022-00009
- Event Type
- Malfunction
- Date Received
- January 24, 2022
- Date of Event
- January 1, 2022
- Report Date
- March 14, 2022
- Manufacturer
- BECTON DICKINSON IND. CIRURGICAS LTDA
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED AS A DEFAULT. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
H6: INVESTIGATION SUMMARY: PHOTOS RECEIVED FOR INVESTIGATION. A DEVICE HISTORY REVIEW WAS PERFORMED FOR REPORTED LOT 1132504, A MAINTENANCE ORDER WAS EVIDENCED THAT COULD POTENTIALLY BE RELATED TO THE INCIDENT. POSSIBLE ROOT CAUSE IS ASSOCIATED WITH THE MISMATCH IN THE ROD RAIL. CORRECTIVE ACTIONS WERE TAKEN IMMEDIATELY TO ADJUST THE ROD RAIL.
IT WAS REPORTED THAT 6 BD PERFUSION 50 ML SYRINGES WITH MOUNTED NEEDLE, OPAQUE HAD A BROKEN PLUNGER ROD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "I AM HEREBY COMMUNICATING ABOUT SOME SYRINGES THAT ARE DEFECTIVE, THEY BREAK WHEN THE PLUNGER IS PULLED."
IT WAS REPORTED THAT 6 BD PERFUSION 50 ML SYRINGES WITH MOUNTED NEEDLE, OPAQUE HAD A BROKEN PLUNGER ROD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "I AM HEREBY COMMUNICATING ABOUT SOME SYRINGES THAT ARE DEFECTIVE, THEY BREAK WHEN THE PLUNGER IS PULLED."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1983163 | BD PERFUSION 50 ML SYRINGE WITH MOUNTED NEEDLE, OPAQUE | PISTON SYRINGE | FMF | BECTON DICKINSON IND. CIRURGICAS LTDA | 1132504 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |