FDA Adverse Event
Malfunction
Summary report: N
PENTAX
MDR report key: 13330867
·
Received January 24, 2022
Report
- Report Number
- 9610877-2022-50146
- Event Type
- Malfunction
- Date Received
- January 24, 2022
- Date of Event
- January 12, 2022
- Report Date
- January 24, 2022
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- EOB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL VNL11-J10 IS AVAILABLE IN THE USA WITH A 510K NUMBER K172156. WE CHECKED THE RETURNED UNIT AND CONFIRMED THE CCD MODULE AS DEFECTIVE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD MODULE. IN ADDITION, WE CONFIRMED THAT THE REMOTE CONTROL BUTTONS PERFORATED, THE INSERTION FLEXIBLE TUBE (IFT) BUCKLED, AND THE U/D KNOB CRACKED ; HOWEVER, THEY ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
Description of Event or Problem · 0
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 670784 | PENTAX | VIDEO NASO PHARYNGO LARYNGOSTROBOSCOPE | EOB | HOYA CORPORATION PENTAX TOKYO OFFICE | VNL-1190STK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |