SIDEWINDER PLATE, RT, 7-HOLE, MEDIUM
Report
- Report Number
- 2031009-2022-00001
- Event Type
- Injury
- Date Received
- January 21, 2022
- Date of Event
- December 5, 2021
- Report Date
- January 21, 2022
- Manufacturer
- TRIMED INC
- Product Code
- HRS
- UDI-DI
- 00842188101458
- PMA / PMN Number
- K060041
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- 003
Narratives
A TRIMED DISTRIBUTOR FROM (B)(6), NOTIFIED TRIMED ON 05 DEC 2021 THAT A SWOR-7M HAD BROKEN AND NEEDED TO BE REMOVED FROM THE PATIENT. THE DATE BELIEVED TO BE THE DATE OF REMOVAL WAS (B)(6) 2021. IT IS CURRENTLY BELIEVED THAT THE BREAK IN THE PLATE WAS CAUSED BY A NONUNION, HOWEVER ADDITIONAL INFORMATION HAS BEEN REQUESTED WITH NO RESPONSE YET. THIS DEVICE WAS IMPLANTED FOR APPROXIMATELY 3 MONTHS BEFORE THE BREAK WAS DETECTED IN AN X-RAY WHICH PROMPTED THE REMOVAL OF THE DEVICE. IT IS CURRENTLY BELIEVED THAT THE PATIENT HAS SINCE STARTED HEALING WELL, HOWEVER ADDITIONAL INFORMATION HAS BEEN REQUESTED WITH NO RESPONSE YET. ADDITIONAL INFORMATION WAS REQUESTED ON THE FOLLOWING DATES, 06 DEC 2021, 09 DEC 2021, 14 DEC 2021, 16 DEC 2021, 28 DEC 2021 AND 03 JAN 2022, WITH NO ADDITIONAL INFORMATION PROVIDED AS OF 04 JAN 2022.
A TRIMED DISTRIBUTOR FROM (B)(6) MEDICAL LOCATED IN CHINA, NOTIFIED TRIMED ON (B)(6) 2022 THAT THE PLATE HAD NOT BROKEN DUE TO A NONUNION BUT MORE LIKELY THE PLATE BROKE DUE TO THE PATIENT NOT FOLLOWING THE CORRECT AFTER CARE INSTRUCTIONS. IT WAS DETERMINED THAT THE PATIENT HAD BEEN PARTICIPATING IN SPORTS WHICH CAUSED THE PATIENT TO FALL AND RE-BREAK THE PLATE AND BONE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 661203 | SIDEWINDER PLATE, RT, 7-HOLE, MEDIUM | SIDEWINDER PLATE, RT, 7-HOLE, MEDIUM | HRS | TRIMED INC | SWOR-7M | MX20007 | 00842188101458 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Required Intervention |