FDA Adverse Event Injury Summary report: N

SIDEWINDER PLATE, RT, 7-HOLE, MEDIUM

MDR report key: 13322318 · Received January 21, 2022

Report

Report Number
2031009-2022-00001
Event Type
Injury
Date Received
January 21, 2022
Date of Event
December 5, 2021
Report Date
January 21, 2022
Manufacturer
TRIMED INC
Product Code
HRS
UDI-DI
00842188101458
PMA / PMN Number
K060041
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
003

Narratives

Description of Event or Problem · 0

A TRIMED DISTRIBUTOR FROM (B)(6), NOTIFIED TRIMED ON 05 DEC 2021 THAT A SWOR-7M HAD BROKEN AND NEEDED TO BE REMOVED FROM THE PATIENT. THE DATE BELIEVED TO BE THE DATE OF REMOVAL WAS (B)(6) 2021. IT IS CURRENTLY BELIEVED THAT THE BREAK IN THE PLATE WAS CAUSED BY A NONUNION, HOWEVER ADDITIONAL INFORMATION HAS BEEN REQUESTED WITH NO RESPONSE YET. THIS DEVICE WAS IMPLANTED FOR APPROXIMATELY 3 MONTHS BEFORE THE BREAK WAS DETECTED IN AN X-RAY WHICH PROMPTED THE REMOVAL OF THE DEVICE. IT IS CURRENTLY BELIEVED THAT THE PATIENT HAS SINCE STARTED HEALING WELL, HOWEVER ADDITIONAL INFORMATION HAS BEEN REQUESTED WITH NO RESPONSE YET. ADDITIONAL INFORMATION WAS REQUESTED ON THE FOLLOWING DATES, 06 DEC 2021, 09 DEC 2021, 14 DEC 2021, 16 DEC 2021, 28 DEC 2021 AND 03 JAN 2022, WITH NO ADDITIONAL INFORMATION PROVIDED AS OF 04 JAN 2022.

Description of Event or Problem · 0

A TRIMED DISTRIBUTOR FROM (B)(6) MEDICAL LOCATED IN CHINA, NOTIFIED TRIMED ON (B)(6) 2022 THAT THE PLATE HAD NOT BROKEN DUE TO A NONUNION BUT MORE LIKELY THE PLATE BROKE DUE TO THE PATIENT NOT FOLLOWING THE CORRECT AFTER CARE INSTRUCTIONS. IT WAS DETERMINED THAT THE PATIENT HAD BEEN PARTICIPATING IN SPORTS WHICH CAUSED THE PATIENT TO FALL AND RE-BREAK THE PLATE AND BONE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
661203 SIDEWINDER PLATE, RT, 7-HOLE, MEDIUM SIDEWINDER PLATE, RT, 7-HOLE, MEDIUM HRS TRIMED INC SWOR-7M MX20007 00842188101458

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention