FDA Adverse Event Malfunction Summary report: N

CADD-LEGACY PLUS PUMP

MDR report key: 13315748 · Received January 21, 2022

Report

Report Number
3012307300-2022-01387
Event Type
Malfunction
Date Received
January 21, 2022
Report Date
January 21, 2022
Manufacturer
ST PAUL
Product Code
FRN
UDI-DI
10610586019647
PMA / PMN Number
K982836
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED AND VISUAL INSPECTION REVEALED HOUSING INTACT, SCREEN SCRATCHED. THE REPORTED EVENT WAS CONFIRMED. A DHR REVIEW WAS PERFORMED SUBSEQUENT TO THE MANUFACTURING OF THE DEVICE AND PRIOR TO ITS RELEASE. NO PROBLEMS OR ISSUES WERE IDENTIFIED DURING THIS DHR REVIEW. THE DEVICE WAS POWERED UP AND ALARMED EC(B)(4) WHICH IS A CORRUPTION OF THE MEMORY OF THE CIRCUIT BOARD. REPLACED CIRCUIT BOARD.

Description of Event or Problem · 0

IT WAS REPORTED THAT (B)(6) PUMP DISPLAYED ERROR CODE DURING TESTING. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
729865 CADD-LEGACY PLUS PUMP PUMP, INFUSION FRN ST PAUL 6500 10610586019647

Patients

Seq Age Sex Outcome Treatment
1 Unknown