MINIMED 780G OUS SYSTEM BLE CONNECT 3.0 MG/DL
Report
- Report Number
- 2032227-2022-104989
- Event Type
- Malfunction
- Date Received
- January 21, 2022
- Date of Event
- January 19, 2022
- Report Date
- June 28, 2022
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000384289
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE PUMP PASSED THE SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, SELF TEST AND DISPLACEMENT TEST. SUCCESSFULLY DOWNLOADED THE TRACE AND HISTORY FILES USING THUS. NO UNEXPECTED LOW BATTERY ALERTS OR POWER RELATED ALARMS NOTED DURING TESTING. A REVIEW OF THE HISTORY FILES REVEALS THAT PUMP ERROR 25 ALARMED ON (B)(6) 2021 (12:00 AND 16:00), ON (B)(6) 2021 (11:00, 15:00, 19:00 AND 23:00), 12/23/2021 (03:00, 0700, 15:00, 19:00, AND 23:00), ON (B)(6) 2021 (03:00, 07:00, 11:00, AND 15:00), AND ON (B)(6) 2022 (23:00). NO UNEXPECTED LOW BATTERY ALERTS NOTED DURING TESTING OR WERE FOUND IN THE HISTORY FILES. PUMP TRACE DOWNLOAD ANALYSIS CONFIRMED THE PUMP ALARMED POWER ERROR 25. THE ADAPT TOOL CONFIRMED POWER ERROR 25 ALARMS WERE TRIGGERED WHEN THE BATTERY LOADED VOLTAGE (LOADED VLITH) WAS LESS THAN 3.5V FOR 4 CONSECUTIVE HOURS DUE CONNECTOR RESISTANCE ON J6 /PCB 1. AFTER DISCONNECTING AND RECONNECTING THE INTERNAL BATTERY CONNECTOR ON J6/PCB 1, THE PUMP WAS MONITORED AND FUNCTIONED PROPERLY. A TEST P-CAP DOES LOCK INTO PLACE INSIDE THE RESERVOIR COMPARTMENT PROPERLY. THE FOLLOWING WERE NOTED DURING PHYSICAL INSPECTION: SCRATCHED CASE, PILLOWING KEYPAD OVERLAY, STAINED KEYPAD OVERLAY, END CAP ADDRESS LABEL FADED, CRACKED BATTERY COMPARTMENT, AND CRACKED BATTERY TUBE THREADS. POWER ERROR 25 ALARMS ARE CONFIRMED DUE TO CONNECTOR RESISTANCE ON J6 /PCB 1. LOW BATTERY LIFE COMPLAINT IS NOT CONFIRMED. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD GENERATED AN ALARM WHEN A POWER SYSTEM ERROR WAS DETECTED AND THIS ERROR DID NOT PREVENT THE INSULIN PUMP FROM RUNNING. CUSTOMER WAS ABLE TO CLEAR THE ALARM AND REWIND ALSO. THE NOTIFICATION LIGHT STOPPED FLASHING IMMEDIATELY. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 943658 | MINIMED 780G OUS SYSTEM BLE CONNECT 3.0 MG/DL | ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1886 | HG4UYEAZZ | 000000763000384289 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |