ALL-IN-ONE
Report
- Report Number
- 3016521623-2022-00009
- Event Type
- Malfunction
- Date Received
- January 20, 2022
- Date of Event
- December 30, 2021
- Report Date
- January 20, 2022
- Manufacturer
- LUCIRA HEALTH, INC.
- Product Code
- QJR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
PRODUCT WAS NOT RETURNED. A DHR REVIEW (DHR SAMPLE VIAL LOT #: 2107044. (INTERNAL LOT#: (200819-8B) AND DHR TEST LOT#: 2106075 (INTERNAL LOT #: 210348-3Y) OF THE ASSOCIATED KIT LOT NUMBER (K07A113107214M2) WAS COMPLETED, AND 0 DISCREPANCIES WERE FOUND. A REVIEW OF EXISTING CAPA, SCAR AND NCMRS WAS COMPLETED AND THERE ARE NO PRIOR RECORDS RELATED TO FALSE POSITIVE FAILURE MODE FOR THIS LOT. FALSE POSITIVE IS A KNOWN POSSIBLE OUTCOME FOR A VERY SMALL PERCENTAGE OF CASES. ROOT CAUSE CANNOT BE DETERMINED BASED ON THE INFORMATION AVAILABLE. POTENTIAL ROOT CAUSES INCLUDE BUT ARE NOT LIMITED TO: ENVIRONMENTAL CONTAMINATION, LOW VIRAL LOAD, AND/OR DEVICE FAILURE. THIS DEVICE IS MARKETED UNDER EUA (B)(4) ALL-IN-ONE.
ONE DEVICE REPORTED AS HAVING A FALSE POSITIVE RESULT. COMPLAINANT PERFORMED AN ADDITIONAL PCR TEST WITH A NEGATIVE RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 484923 | ALL-IN-ONE | COVID-19 TEST KIT | QJR | LUCIRA HEALTH, INC. | K07A113107214M2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |