FDA Adverse Event Malfunction Summary report: N

GAMMA LNG NL LFT 11X360MM 130

MDR report key: 133072 · Received November 14, 1997

Report

Report Number
9610622-1997-00127
Event Type
Malfunction
Date Received
November 14, 1997
Date of Event
March 17, 1997
Report Date
November 12, 1997
Manufacturer
HOWMEDICA INC.
Product Code
HSB
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

SUMMARY OF EVAL: EVAL RESULTS SUGGEST THAT THE NAIL BROKE DUE TO MATERIAL FATIGUE.

Description of Event or Problem · 1

OCTOBER 21, 1995, A GAMMA NAIL WAS IMPLANTED FOR A BROKEN FEMUR. 3-17-1997, REVISION SURGERY WAS REQUIRED BECAUSE OF THE FEMUR'S BREAKAGE ON PSEUDOARTHROSIS. THE BROKEN GAMMA NAIL WAS REPLACED BY A PLATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GAMMA LNG NL LFT 11X360MM 130 Implant IMPLANT HSB HOWMEDICA INC. NA K320844

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other GAMMA LAG SCREW, 12X90 MM