FDA Adverse Event Malfunction Summary report: N

PROLENE SUTURE 36"(90CM) 5-0 BLU

MDR report key: 13283222 · Received January 18, 2022

Report

Report Number
2210968-2022-00467
Event Type
Malfunction
Date Received
January 18, 2022
Date of Event
December 23, 2021
Report Date
March 23, 2022
Manufacturer
ETHICON INC.
Product Code
GAW
UDI-DI
10705031022058
PMA / PMN Number
K133356
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MK
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). DATE SENT TO THE FDA: 3/23/2022. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE BATCH AND AN NC [NR-0141286] RELATED TO THE REPORTED INCIDENT WAS IDENTIFIED. THE NECESSARY ACTIONS TO ENSURE THE FINAL PRODUCT QUALITY HAVE BEEN TAKEN AND DOCUMENTED IN THE APPROPRIATE QUALITY SYSTEM. THE FINAL QUALITY RELEASE CRITERIA WERE MET BEFORE THIS BATCH WAS RELEASED FOR DISTRIBUTION. H6 COMPONENT CODE: G07002 NO DEVICE PROBLEM FOUND. H3 INVESTIGATIONAL NARRATIVE: FIFTY PACKETS OF PRODUCT CODE 8934H WERE RETURNED TO ETHICON INC FOR ANALYSIS WITH THE PACKAGING CLOSED. UPON INITIAL INSPECTION, TO THE SAMPLES, NO EXTERNAL DAMAGES WERE OBSERVED ON THE PACKETS. IN ORDER TO EVALUATE THE CONDITIONS OF THE RETURNED SAMPLES, THIRTEEN PACKETS WERE OPENED, AND NO DEFECTS WERE DETECTED. THE SWAGE AND ATTACHMENT AREA WERE NOTED TO BE AS EXPECTED. THE SUTURES WERE DISPENSED WITHOUT PROBLEMS AND EXAMINED ALONG THE STRAND TO DETECT ANY ISSUE RELATED TO DAMAGED OR BREAKAGE SUTURE AND NO DEFECTS WERE OBSERVED DURING EVALUATION. A FUNCTIONAL TEST WAS PERFORMED USING AN INSTRON AND THE TENSILE STRENGTH FORCE WAS ABOVE THE MINIMUM REQUIREMENT. THE EVENT DESCRIBED COULD NOT BE CONFIRMED AS THE DEVICE PERFORMED WITHOUT ANY DEFECT NOTED. AS PART OF OUR QUALITY PROCESS, THE MANUFACTURING RECORDS OF THIS LOT-SERIAL NUMBER WERE REVIEWED, AND THE MANUFACTURING STANDARDS WERE MET PRIOR TO THE RELEASE OF THIS BATCH. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH BATCH IS RANDOMLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS IS CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST-MARKET SURVEILLANCE. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON INC, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON INC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

(B)(4). ATTEMPTS HAVE BEEN MADE TO RETRIEVE THE DEVICE. TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. ATTEMPTS ARE BEING MADE TO OBTAIN THE FOLLOWING INFORMATION. TO DATE NO RESPONSE HAS BEEN PROVIDED. IF FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. WHAT TISSUE WAS BEING APPROXIMATED OR SUTURED WHEN IT BROKE? WHAT WAS USED TO COMPLETE THE PROCEDURE? EVENTS REPORTED IN 2210968-2022-00464, 2210968-2022-00465, 2210968-2022-00466 AND 2210968-2022-00468.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). DATE SENT TO THE FDA: 02/15/2022. H6 COMPONENT CODE: G07002 - DEVICE NOT RETURNED. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON INC, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON INC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE, AND NO NON-CONFORMANCES WERE IDENTIFIED. ADDITIONAL INFORMATION WAS REQUESTED AND THE FOLLOWING WAS OBTAINED: NO RESPONSE WAS RECEIVED FOR THE TISSUE-RELATED QUESTION, THE SALESPERSON WAS TOLD THEY HAVE FINISHED THE OPERATION WITH SURGIPRO (MEDTRONIC). WHAT TISSUE WAS BEING APPROXIMATED OR SUTURED WHEN IT BROKE? WHAT WAS USED TO COMPLETE THE PROCEDURE?

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN CARDIOVASCULAR PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
162926 PROLENE SUTURE 36"(90CM) 5-0 BLU SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE GAW ETHICON INC. 8934H QBBLPZ 10705031022058

Patients

Seq Age Sex Outcome Treatment
1 Unknown