FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 13262988 · Received January 16, 2022

Report

Report Number
9610877-2022-00143
Event Type
Malfunction
Date Received
January 16, 2022
Date of Event
January 12, 2022
Report Date
March 11, 2022
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
FDS
UDI-DI
04961333237586
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EG29-I10C-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K190805. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, A SUPPLEMENTAL REPORT WILL BE FILED WITH THE NEW INFORMATION.

Additional Manufacturer Narrative · 0

EVALUATION SUMMARY: THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EG29-I10C-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K190805. WE CHECKED THE RETURNED UNIT AND CONFIRMED THE CCD MODULE AS DEFECTIVE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD MODULE. IN ADDITION, WE CONFIRMED THAT THE LG CABLE CONNECTOR HOUSING LEAKAGE; HOWEVER, IT IS NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR. CORRECTION INFORMATION: G6: FOLLOW UP #1 H6: CODING CHANGED BASED ON THE INVESTIGATION RESULT ADDITIONAL INFORMATION: H4: DEVICE MANUFACTURE DATE.

Description of Event or Problem · 0

FOGGY IMAGE. MOISTURE INSIDE. A/W SOCKET LOOSENED. THIS EVENT OCCURRED AT THE TIME OF DURING INSPECTION. THERE WAS NO REPORT OF PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
701657 PENTAX IMAGINA GASTROSCOPE FDS HOYA CORPORATION PENTAX TOKYO OFFICE EG29-I10C 04961333237586

Patients

Seq Age Sex Outcome Treatment
1 Unknown