FOLEY CATHETER, ADULT 400 SERIES, 16 FR.
Report
- Report Number
- 1221261-2008-00039
- Event Type
- Malfunction
- Date Received
- December 18, 2008
- Date of Event
- August 8, 2008
- Report Date
- August 22, 2008
- Manufacturer
- MRI MEDICAL INC.
- Product Code
- KOD
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
EVALUATION - OTHER: INVESTIGATION: THE RETURNED EVENT SAMPLE WAS FORWARDED TO THE SUPPLIER, MRI MEDICAL. A VISUAL INSPECTION OF THE CATHETER SHOWED A TEAR PATTERN IN THE SHAFT SUCH THAT THE CATHETER APPEARED TO BE UNDER TENSILE STRESS PRIOR TO TEARING. A REVIEW OF ALL MANUFACTURING RECORDS DID NOT IDENTIFY ANY RELATED NON CONFORMANCES OR OTHER ISSUES RELATED TO THE SHAFT QUALITY. THE QUALITY CONTROL INSPECTION FOR THIS PRODUCT INCLUDES INSPECTION FOR THIS TYPE OF DEFECT WITH THE REQUIREMENT THAT "SHAFT IS SMOOTH AND FREE OF KINKS, CUTS, TEARS, IRREGULAR SURFACE OR CONTAMINATION." ROOT CAUSE: THE MOST LIKELY CAUSE FOR TEARING OF THE SHAFT IS THAT A CUT OR NICK WAS PRESENT IN THE SILICONE. A CUT OR NICK PRESENT IN THE SILICONE MAY HAVE BEEN PROPOGATED WITH ADDITIONAL HANDLING AND USE; THIS DEFECT MAY HAVE PROPOGATED TO THE POINT OF FAILURE DUE TO STRAIN WHEN IN USE BY THE PATIENT OR BY THE PATIENT TAMPERING WITH THE DEVICE. WE BELIEVE THIS TO BE AN ISOLATED EVENT LIMITED TO THE SAMPLE RECEIVED AS THERE IS NO HISTORY OF DEFECTS OF THIS NATURE FOR THIS PRODUCT. THIS REPORT HAS BEEN LOGGED FOR TRENDING.
USER REPORTS THAT A REMNANT PART OF THE TUBING BROKE OFF AND HAD TO BE SURGICALLY REMOVED FROM THE PATIENT. THE CATHETER HAD BEEN IN PLACE FOR ONE WEEK. NO ADVERSE OUTCOME REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FOLEY CATHETER, ADULT 400 SERIES, 16 FR. | 78 K - - OD CATHETER, UROLOGICAL, SELF-RETAINED | KOD | MRI MEDICAL INC. | 687302 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | Required Intervention |