FDA Adverse Event Injury Summary report: N

UNKNOWN ABGII STEM

MDR report key: 13257061 · Received January 14, 2022

Report

Report Number
0002249697-2022-00081
Event Type
Injury
Date Received
January 14, 2022
Date of Event
October 24, 2020
Report Date
May 30, 2022
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
MEH
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

G1 MANUFACTURING SITE CORRECT TO SHC. AN EVENT REGARDING A CRACK/FRACTURE INVOLVING AN UNKNOWN ABG II STEM WAS REPORTED. THE EVENT WAS NOT CONFIRMED. METHOD & RESULTS: -DEVICE EVALUATION AND RESULTS: NOT PERFORMED AS PRODUCT WAS NOT RETURNED. -CLINICIAN REVIEW: NO MEDICAL RECORDS WERE RECEIVED FOR REVIEW WITH A CLINICAL CONSULTANT -DEVICE HISTORY REVIEW: COULD NOT BE PERFORMED AS LOT CODE INFORMATION WAS NOT PROVIDED. -COMPLAINT HISTORY REVIEW: COULD NOT BE PERFORMED AS LOT CODE INFORMATION WAS NOT PROVIDED. CONCLUSION: THE EXACT CAUSE OF THE EVENT COULD NOT BE DETERMINED BECAUSE INSUFFICIENT INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION INCLUDING OPERATIVE REPORTS, PROGRESS NOTES, X-RAYS AND RETURN OF THE DEVICE ARE NEEDED TO FULLY INVESTIGATE THE EVENT. IF FURTHER INFORMATION BECOMES AVAILABLE OR THE PRODUCT IS RETURNED, THIS INVESTIGATION WILL BE RE-OPENED.

Additional Manufacturer Narrative · 0

IT WAS NOTED THAT THE DEVICE IS NOT AVAILABLE FOR EVALUATION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT UPON COMPLETION OF THE INVESTIGATION. DEVICE NOT RETURNED TO THE MANUFACTURER.

Description of Event or Problem · 0

PATIENT PRESENTED TO HOSPITAL WITH LEFT LATERAL THIGH PAIN, WHERE IT WAS FOUND THERE HAD BEEN A SECOND TRANSVERSE FRACTURE THROUGH THE PROSTHESIS OF THE STEM OF THE LEFT HIP.

Description of Event or Problem · 0

PATIENT PRESENTED TO HOSPITAL WITH LEFT LATERAL THIGH PAIN, WHERE IT WAS FOUND THERE HAD BEEN A SECOND TRANSVERSE FRACTURE THROUGH THE PROSTHESIS OF THE STEM OF THE LEFT HIP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
682973 UNKNOWN ABGII STEM PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, NON-POROUS, CALICU MEH STRYKER ORTHOPAEDICS-MAHWAH UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 73 YR Male Hospitalization| R