FDA Adverse Event Malfunction Summary report: N

BINAXNOW¿ COVID-19 ANTIGEN SELF TEST

MDR report key: 13252561 · Received January 13, 2022

Report

Report Number
1221359-2022-00227
Event Type
Malfunction
Date Received
January 13, 2022
Date of Event
December 23, 2021
Report Date
June 10, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA210264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. REFERENCE MFR REPORTS: 1221359-2022-00226.

Additional Manufacturer Narrative · 0

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 174799 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160/ LOT 174799 AND TEST BASE PART NUMBER 195-430H / LOT 171888. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 174799 SHOWED THAT THE COMPLAINT RATE IS(B)(4). IN CONCLUSION, ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD POSSIBLY BE RELATED TO ISSUES INCLUDING THE SELF-TEST USER PERFORMANCE, INTERPRETATION OF THE RESULT, OR THE SPECIFIC PATIENT SAMPLE.

Description of Event or Problem · 0

THE CONSUMER REPORTED TWO (2) FALSE POSITIVE RESULTS WITH THE BINAXNOW¿ COVID-19 ANTIGEN SELF TEST FOR A PATIENT PERFORMED IN BETWEEN (B)(6) 2021 . THIS MFR. REPORT ADDRESSES DATE (B)(6) 2021 AND IS TWO (2) OF TWO (2). THE CONSUMER REPORTED A FALSE POSITIVE RESULT WITH THE BINAXNOW¿ COVID-19 ANTIGEN SELF TEST ON (B)(6) 2021 TESTED ON A NASAL KITTED SWAB. ON (B)(6) 2021, PCR CONFIRMATION TESTING GENERATED A NEGATIVE RESULT. THE CONSUMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO THE TEST RESULTS. ADDITIONALLY, THE CUSTOMER CONFIRMED THERE WAS NO DELAY OR IMPACT IN THE PATIENT'S TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
777932 BINAXNOW¿ COVID-19 ANTIGEN SELF TEST LATERAL FLOW IMMUNOASSAY IVD OF COVID-19 QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 174799 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 31 YR Female