- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE PRODUCT HAS NOT BEEN RETURNED TO THE MFR FOR EVAL. WE ARE UNABLE TO CONFIRM ANY PRODUCT MALFUNCTION. NO CONCLUSION CAN BE REACHED AT THIS TIME. THE CUSTOMER STATED THAT A KINK WAS NOTED IN THE CANNULA. THERE WAS NO REPORT OF AN ALARM, HOWEVER, THE POD WOULD HAVE INITIATED AN OCCLUSION ALARM IF THE FLOW OF INSULIN WAS RESTRICTED/PREVENTED BY THE KINK, AND RESULTED IN BACK PRESSURE. THE PRODUCT USER GUIDE INSTRUCTS THE USER TO "CHECK INFUSION SITE FREQUENTLY FOR PROPER CANNULA PLACEMENT". IT ALSO SUGGESTS THAT THE USER SHOULD CHECK THEIR BLOOD GLUCOSE LEVELS FREQUENTLY, SO THAT THEY CAN NOTICE AND REACT QUICKLY AND APPROPRIATELY TO ANY ISSUES. THE PT REMEDIED THE SITUATION BY REMOVING AND REPLACING THE DEVICE PER USER INSTRUCTIONS.
CUSTOMER CALLED TO REPORT A POD THAT SHE FELT WAS NOT DELIVERING INSULIN. UPON REMOVAL OF THE POD, SHE SAID THE CANNULA APPEARED KINKED. HER BLOOD GLUCOSE LEVEL AT THE TIME WAS 300 MG/DL. NO FURTHER ISSUES WERE IDENTIFIED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OMNIPOD INSULIN PUMP | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 11200 | L11814 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 25 YR | Other |