FDA Adverse Event Malfunction Summary report: N

HEADBAND HB-7 FOR TDH39

MDR report key: 13235218 · Received January 12, 2022

Report

Report Number
9612197-2022-00001
Event Type
Malfunction
Date Received
January 12, 2022
Date of Event
December 22, 2021
Report Date
February 10, 2022
Manufacturer
NATUS MEDICAL DENMARK APS
Product Code
EWO
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

INITIAL REPORT REFERENCE TO COMPLAINT#: (B)(4). IT WAS ESTABLISHED THAT THE HEADPHONES HAD NOT BEEN REPLACED AND ARE THEREFORE LIKELY IN SERVICE SINCE INSTALLED WITH ORIGINAL DEVICE ((B)(6) 2015) THE SUPPLIER HAS CHANGED THE MATERIAL USED IN THE MANUFACTURE OF THE HEADBAND SINCE THESE HEADPHONES WERE INSTALLED. THE CURRENT RISK FILE (B)(4) REV 03 - 1081 AURICAL AUD & MADSEN A450 - RISK ANALYSIS HAZARD ID 6.5 IDENTIFIES THIS ISSUE HARM - MINOR INJURY TO THE PATIENT OR USER OR DISCOMFORT. CAUSE- MECHANICAL BREAKAGE, CAUSING SHARP CORNER, EDGES OR PINCH POINTS. E.G. DUE TO STEADY FORCE APPLIED ON THE DEVICE. (HIT BY SOMETHING) SEVERITY- MARGINAL (3), RISK LEVEL- MINOR (3). A RISK REVIEW IS NOT REQUIRED AS THIS COMPLAINT DOES NOT DESCRIBE A NEW FAILURE MODE OR NEW HARM AND THE EXISTING HAZARD SEVERITY AND/OR PROBABILITY OF OCCURRENCE HAS NOT CHANGED. A SEARCH FOR SERVICE DATA THAT MAY BE RELEVANT TO THIS ISSUE HAS BEEN CONDUCTED. NO FURTHER INFORMATION RELATED TO THIS DEVICE FOUND. UDI NUMBER: NOT APPLICABLE. INVESTIGATION STILL ONGOING.

Additional Manufacturer Narrative · 0

FOLLOW UP REPORT 001 REFERENCE TO COMPLAINT# (B)(4). THIS FAILURE WAS PREVIOUSLY INVESTIGATED REFERENCE TO DOC-050766 REV A- COMPLAINT INVESTIGATION REPORT . THE REPORT DOCUMENTS THE FOLLOWING: INVESTIGATION RESULTS: FORCE-TEST. MOUNTED THE NONE-BROKEN SIDE OF THE HEADSET TO A FIXED FIXTURE AND FORCE GAUGES CONNECTED TO THE SIDE WITH THE CRACKED PLASTIC. PRESSURE IS SLOWLY APPLIED PULLING THE FORCE GAUGE STRETCHING THE HEADSET FROM ITS NORMAL POSITION 1 TO POSITION 2. IT WAS POSSIBLE TO STRETCH THE HEADSET 23-24 CM BEFORE THE EAR HOOK BROKE OFF THE HEADBAND AS THE CUSTOMER EXPLAINED. THE FORCE GAUGE SHOWED 14 NEWTON'S AT THE MAXIMUM POSITION, CONVERTING THE VALUE TO KG AND WE GET A VALUE OF 1.4276 KILOGRAM-FORCE. THIS SHOWS THAT A GOOD AMOUNT OF FORCE IS NEEDED TO BREAK OFF THE EAR HOOK, WHEN THE PLASTIC IS BROKEN. THE CUSTOMER EXPLAINS THAT THE HEADSET EAR HOOK WAS AT THE MAXIMUM EXTENDED POSITION WHEN IT BROKE. FORCE-TEST OTHER SIDE. PERFORMED THE FORCE TEST WITH THE SAME HEADSET, THIS TIME THE BROKEN SIDE WAS MOUNTED TO THE FIXED FIXTURE AND FORCE GAUGES CONNECTED TO THE NONE-BROKEN SIDE. PRESSURE IS SLOWLY APPLIED PULLING THE FORCE GAUGE STRETCHING THE HEADSET AS MUCH AS POSSIBLE, HERE IT WAS NOT POSSIBLE TO BREAK THE EAR HOOK OFF THE HEADBAND. THE FORCE-GAUGE SHOWED 90 NEWTON, CONVERTING THE VALUE TO KG AND WE GET A VALUE OF 9.1774 KILOGRAM-FORCE, THIS SHOWS THAT A SUBSTANTIAL AMOUNT OF FORCE IS APPLIED WITHOUT CRACKING THE PLASTIC. WE INVESTIGATED OTHER METHODS OF BREAKING THE PLASTIC SURROUNDING THE EAR HOOK, THE TEST THAT SHOWED THE SAME DAMAGE AS THE CUSTOMERS WAS BY TWISTING THE EAR HOOK UNTIL THE PLASTIC BROKE RELEASING THE EAR HOOK FROM ITS GRIP AND WHERE THE PLASTICS BROKEN, WHEN COMPARING IT WITH THE CUSTOMERS ITS ALMOST IDENTICAL, A LONG TEAR ALONG THE INNER SIDE OF THE PLASTICS OBSERVED. WE ALSO NOTICED THAT THE PLASTIC GRIP AROUND THE EAR HOOK IS LOOSENED CAUSING THE EAR HOOK TO SLIDE EASILY UP AND DOWN AND THIS IS NOT THE CASE WHEN THE PLASTICS UNBROKEN. CONTACTED THE HEADSET MANUFACTURING COMPANY WILFAN IN ORDER TO INVESTIGATE: IF THE USE OF ALCOHOL DISINFECTANT WOULD MAKE THE PLASTIC BRITTLE AROUND THE EAR HOOK. WHAT THE PRODUCT LIFESPAN IS FOR THE HEADBAND. IF THE PLASTIC MATERIAL HAS CHANGED OVER TIME. THE FOLLOWING REPLIES WERE RECEIVED: THE PLASTIC IS POM, ALSO CALLED ACETAL. THE ENGINEER/ DESIGNER MENTIONED THAT ACCORDING TO THE CHEMICAL COMPATIBILITY CHART, ALCOHOL SHOULD NOT AFFECT THE PLASTIC. PRODUCT LIFESPAN - A CYCLE TESTING WAS NEVER DONE. ENGINEERING WILL ADD IT TO THE VALIDATION OF THE NEW PROCESS. THEY NOTED THAT THE ITEM HAS A 1-YEAR WARRANTY SINCE DAY OF PURCHASE. AS FAR AS THEY HAVE BEEN ABLE TO VERIFY, THE MATERIAL HAS NOT CHANGED. HOWEVER, THEY MANUFACTURE ITEMS OUTSIDE OF HEADQUARTERS AND VERIFICATION HAS NOT BEEN DONE BY SOMEONE IN HEADQUARTERS YET. THE ROOT CAUSE OF THE HEADSET BREAKING OFF IS THAT THE PLASTIC SURROUNDING THE EAR HOOK HAS CRACKED AND THIS CAUSES THE PLASTIC TO LOSE ITS STRENGTH TO HOLD ON THE EAR HOOK, WHEN PRESSURE IS APPLIED THE PLASTIC EXPANDS AND ALLOWS THE EAR HOOK TO SLIP OUT OF ITS GRIP. THE INVESTIGATION ALSO SHOWS THAT THE HEADBAND CAN WITHSTAND A BIG AMOUNT OF FORCE WITHOUT BREAKING THE EAR HOOK OFF WHEN THE PLASTIC IS INTACT AND USED IN THE CORRECT WAY, BUT THE INVESTIGATION ALSO SHOW THAT IT'S POSSIBLE TO BREAK THE HEADSET IF IT'S USED INCORRECTLY FOR EXAMPLE TWISTING THE EAR HOOK. RECOMMENDED ACTION: UPDATE THE USER GUIDE, TO INCLUDE A SENTENCE STATING THAT THE ACOUSTICIAN SHOULD INSPECT THE HEADSET FOR ANY DAMAGE BEFORE USING IT ON A PATIENT, TO AVOID USING THE HEADSET WITH A BROKEN HEADSET. COMPLAINTS ARE REVIEWED ROUTINELY PER QUALITY SYSTEM REQUIREMENTS (QMS-004442 CORPORATE TRENDING AND ANALYSIS PROCEDURE) AND COMPLAINT TRENDS ARE ASSESSED AS PART OF THESE REVIEWS. A REVIEW OF COMPLAINT TRENDING IS COMPLETED QUARTERLY.

Description of Event or Problem · 0

HEADBAND HB-7 FOR TDH39 SNAPPED - PART 1-25-12200. THE PLASTIC SECTION OF THE HEADBAND BROKE, CAUSING IT TO BE RELEASED FROM THE METAL ROD ON THE EARPIECE CLAMP. THIS ALLOWED THE HEADBAND TO SNAP BACK AGAINST THE PATIENT. THERE WAS AN ABRASION AND SLIGHT BLEEDING.

Description of Event or Problem · 0

HEADBAND HB-7 FOR TDH39 SNAPPED - PART 1-25-12200. THE PLASTIC SECTION OF THE HEADBAND BROKE, CAUSING IT TO BE RELEASED FROM THE METAL ROD ON THE EARPIECE CLAMP. THIS ALLOWED THE HEADBAND TO SNAP BACK AGAINST THE PATIENT. THERE WAS AN ABRASION AND SLIGHT BLEEDING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
732058 HEADBAND HB-7 FOR TDH39 HEADBAND HB-7 FOR TDH39 EWO NATUS MEDICAL DENMARK APS 1-25-12200

Patients

Seq Age Sex Outcome Treatment
1 0 YR Prefer Not To Disclose Other