FDA Adverse Event Malfunction Summary report: N

BD PEGASUS¿ SAFETY CLOSED IV CATHETER

MDR report key: 13231453 · Received January 11, 2022

Report

Report Number
3014704491-2021-00431
Event Type
Malfunction
Date Received
January 11, 2022
Date of Event
December 9, 2021
Report Date
July 22, 2024
Manufacturer
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
Product Code
FOZ
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

IN RESPONSE TO THE EVENT REPORTED BY YOUR FACILITY A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 1221451. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, VISUAL ANALYSIS WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RETAINED DEVICES WERE FOUND TO BE FREE OF ANY DAMAGE OR OTHER ABNORMALITIES. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT. BD WILL CONTINUE TO MONITOR THIS ISSUE AND ENCOURAGES YOU TO SUBMIT YOUR SAMPLE FOR REVIEW.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING TRANSFUSION WITH BD PEGASUS¿ SAFETY CLOSED IV CATHETER THE CATHETER WAS DISCOVERED TO BE BENT. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: ON THE MORNING OF (B)(6) 2021, NURSING STAFF OF THE RESPIRATORY DEPARTMENT FOUND THAT THE INDWELLING NEEDLE CATHETER WAS BENT WHEN THEY WERE PROVIDING TRANSFUSION FOR THE PATIENT, WHICH MADE IT IMPOSSIBLE TO ADMINISTER MEDICINE NORMALLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
635934 BD PEGASUS¿ SAFETY CLOSED IV CATHETER INTRAVASCULAR CATHETER FOZ BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU 1221451

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown