FDA Adverse Event
Malfunction
Summary report: N
T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY
MDR report key: 13230866
·
Received January 11, 2022
Report
- Report Number
- 3013756811-2022-00046
- Event Type
- Malfunction
- Date Received
- January 11, 2022
- Date of Event
- December 21, 2021
- Report Date
- January 11, 2022
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- OZO
- UDI-DI
- 00853052007264
- PMA / PMN Number
- K201214
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ME, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
Additional Manufacturer Narrative · 0
THE DEVICE IS EXPECTED TO BE RETURNED; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF THE DEVICE IS RECEIVED.
Additional Manufacturer Narrative · 0
THE FAILURE INVESTIGATION HAS BEEN COMPLETED. BASED ON THE ANALYSIS, THE ALLEGED ISSUES COULD NOT BE VERIFIED; HOWEVER, A DIFFERENT ISSUE WAS IDENTIFIED (DAMAGED USB PINS).
Description of Event or Problem · 0
IT WAS REPORTED THAT THE PUMP TOUCH SCREEN WAS INTERMITTENTLY UNRESPONSIVE. DURING TROUBLESHOOTING, THE ISSUE WAS CONFIRMED AND THE PUMP WAS REPLACED. ADDITIONALLY, IT WAS REPORTED THAT AN UNSPECIFIED MALFUNCTION ALARM OCCURRED. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 233-397 MG/DL. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 686633 | T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY | AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND | OZO | TANDEM DIABETES CARE | 1000096 | 00853052007264 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR | Female |