FDA Adverse Event Malfunction Summary report: N

CS100 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V

MDR report key: 13224337 · Received January 11, 2022

Report

Report Number
2249723-2022-00052
Event Type
Malfunction
Date Received
January 11, 2022
Date of Event
December 16, 2021
Report Date
June 29, 2023
Manufacturer
DATASCOPE CORP. - MAHWAH
Product Code
DSP
UDI-DI
10607567107295
PMA / PMN Number
K031636
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

TESTING OF ACTUAL/SUSPECTED DEVICE (10/3233): A GETINGE FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE IABP UNIT. THE FSE CHECKED IF THE POWER ON THE SCREEN HAD WHITE STRIPES. THE FSE CHANGED THE BOARD PCB, COLOR VIDEO RECEIVER WITH A NEW BOARD. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON RECEIPT OF ADDITIONAL INFORMATION.

Additional Manufacturer Narrative · 0

TESTING OF ACTUAL/SUSPECTED DEVICE (10/3233): A GETINGE FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE IABP UNIT. THE FSE CHECKED IF THE POWER ON THE SCREEN HAD WHITE STRIPES. THE FSE REPLACED THE COLOR VIDEO RECEIVER BOARD AND NOTED A PIECE OF CABLE WAS STILL PENDING DELIVERY IN ORDER TO COMPLETE REPAIRS. SUBSEQUENTLY, THE FSE RETURNED TO THE CUSTOMER'S SITE AT A LATER DATE AND REPLACED THE VIDEO RECEIVER TO LCD CABLE. ALL SAFETY, FUNCTIONALITY AND CALIBRATION TESTS WERE PERFORMED AND PASSED AND THE IABP UNIT WAS THEN CLEARED FOR USE AND RETURNED TO THE CUSTOMER. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON RECEIPT OF ADDITIONAL INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING ROUTINE CHECK THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) HAD A SCREEN THAT WAS DISPLAYING WHITE LINES AND FLICKERED. THERE WAS NO PATIENT INVOLVEMENT, AND NO ADVERSE EVENT WAS REPORTED.

Description of Event or Problem · 0

N/A

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
688155 CS100 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL DSP DATASCOPE CORP. - MAHWAH 0998-00-3013-55 N/A 10607567107295

Patients

Seq Age Sex Outcome Treatment
1 Unknown