BD RAPID DETECTION OF SARS-COV-2 VERITOR¿
Report
- Report Number
- 1119779-2021-02123
- Event Type
- Malfunction
- Date Received
- January 10, 2022
- Date of Event
- October 20, 2021
- Report Date
- June 7, 2022
- Manufacturer
- BECTON, DICKINSON & CO. (SPARKS)
- Product Code
- QKP
- UDI-DI
- 00382902560661
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
EUA # (B)(4). A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H.6. INVESTIGATION: THIS STATEMENT IS TO SUMMARIZE THE INVESTIGATION RESULTS REGARDING THE COMPLAINTS THAT ALLEGES FALSE NEGATIVE RESULTS WHEN USING BD KIT VERITOR FOR RAPID DETECTION OF SARS-COV-2 (MN# 256082), BATCH NUMBER 1153341. BD QUALITY PERFORMS A SYSTEMATIC APPROACH TO INVESTIGATE FALSE NEGATIVE COMPLAINTS. THIS APPROACH INVOLVES REVIEW OF MANUFACTURING BATCH HISTORY RECORDS, TESTING OF RETENTION SAMPLES, AND TESTING OF CUSTOMER RETURNED SAMPLES IF APPLICABLE. A BATCH REVIEW WAS PERFORMED FOR THE BATCH NUMBER PROVIDED. A CONVERSATION WAS CONDUCTED BETWEEN THE CUSTOMER AND BD MEDICAL AFFAIRS ABOUT SENSITIVITY LEVELS AND TESTING OF ASYMPTOMATIC PATIENTS. THE CUSTOMER UNDERSTOOD THAT CLINICAL EVIDENCE SUGGESTS THAT THERE WILL BE LOWER SENSITIVITY AMONG ASYMPTOMATIC PEOPLE. THE CUSTOMER UNDERSTOOD THE ISSUE, BUT WILL BE PROCEEDING WITH AN OFF LABEL VERIFICATION INVOLVING ASYMPTOMATIC PATIENTS. THE REPORTED ISSUE WAS UNABLE TO BE CONFIRMED. THE RETENTION TESTING COULD NOT BE TESTED AS THEY ARE EXPIRED. THE ROOT CAUSE COULD NOT BE IDENTIFIED. NO TREND AGAINST FALSE NEGATIVE WAS IDENTIFIED. CURRENTLY NO ADVERSE TREND FOR FALSE NEGATIVE WAS IDENTIFIED. BD QUALITY WILL CONTINUE TO CLOSELY MONITOR FOR TRENDS. SEE H.10.
IT WAS REPORTED THAT WHILE TESTING WITH THE BD RAPID DETECTION OF SARS-COV-2 VERITOR¿, A FALSE NEGATIVE RESULT WAS OBTAINED. CONFIRMATORY PCR TESTING WAS PERFORMED, AND THE RESULTS WERE POSITIVE. THERE WAS NO REPORT OF PATIENT IMPACT. EUA # (B)(4). THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: SAMPLE NO. (B)(6) WAS TESTED BY RT-PCR FOR SARS-COV-2 (NASOPHARYNX AND OROPHARYNX SWAB IN 0.9% SALINE SOLUTION), WITH A POSITIVE RESULT AND CTS OF 24.6 FOR THE N2 GENE AND 21.8 FOR THE GENE E (GENEXPERT). IT WAS THEN TESTED ON THE BD VERITOR SARS-COV-2, WITH A NEGATIVE RESULT.
IT WAS REPORTED THAT WHILE TESTING WITH THE BD RAPID DETECTION OF SARS-COV-2 VERITOR¿, A FALSE NEGATIVE RESULT WAS OBTAINED. CONFIRMATORY PCR TESTING WAS PERFORMED, AND THE RESULTS WERE POSITIVE. THERE WAS NO REPORT OF PATIENT IMPACT. (B)(6) THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: SAMPLE NO. 2129315316 WAS TESTED BY RT-PCR FOR SARS-COV-2 (NASOPHARYNX AND OROPHARYNX SWAB IN 0.9% SALINE SOLUTION), WITH A POSITIVE RESULT AND CTS OF 24.6 FOR THE N2 GENE AND 21.8 FOR THE THE GENE E (GENEXPERT). IT WAS THEN TESTED ON THE BD VERITOR SARS-COV-2, WITH A NEGATIVE RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1443934 | BD RAPID DETECTION OF SARS-COV-2 VERITOR¿ | CORONAVIRUS ANTIGEN DETECTION SYSTEM | QKP | BECTON, DICKINSON & CO. (SPARKS) | 256066 | 1153341 | 00382902560661 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |