FDA Adverse Event Other Summary report: N

OMNIPOD INSULIN PUMP

MDR report key: 1321437 · Received February 13, 2009

Report

Report Number
3004464228-2009-00098
Event Type
Other
Date Received
February 13, 2009
Date of Event
January 17, 2009
Report Date
January 17, 2009
Manufacturer
INSULET CORPORATION
Product Code
LZG
PMA / PMN Number
K042792
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT HAS NOT BEEN RETURNED TO THE MFR FOR EVALUATION. WE ARE UNABLE TO CONFIRM ANY PRODUCT MALFUNCTION. NO CONCLUSION CAN BE REACHED AT THIS TIME. THE CUSTOMER STATED THAT A KINK WAS NOTED IN THE CANNULA. THERE WAS NO REPORT OF AN ALARM, HOWEVER, THE POD WOULD HAVE INITIATED AN OCCLUSION ALARM IF THE FLOW OF INSULIN WAS RESTRICTED/PREVENTED BY THE KINK AND RESULTED IN BACK PRESSURE. THE PRODUCT USER GUIDE INSTRUCTS THE USER TO "CHECK INFUSION SITE FREQUENTLY FOR PROPER CANNULA PLACEMENT". IT ALSO SUGGESTS THAT THE USER SHOULD CHECK THEIR BLOOD GLUCOSE LEVELS FREQUENTLY, SO THAT THEY CAN NOTICE AND REACT QUICKLY AND APPROPRIATELY TO ANY ISSUES. THE PT REMEDIED THE SITUATION BY REMOVING AND REPLACING THE DEVICE PER USER INSTRUCTIONS.

Description of Event or Problem · 1

CUSTOMER'S DAD CALLED TO REPORT THAT HE DID NOT THINK HIS SON'S POD WAS WORKING PROPERLY. CUSTOMER'S FATHER STATED THE CANNULA WAS BENT WHEN THE POD WAS REMOVED. CUSTOMER WAS ABLE TO ACTIVATE A NEW POD. CUSTOMER'S BLOOD GLUCOSE LEVELS WERE ELEVATED AS HIGH AS 500 MG/DL WHILE WEARING THIS POD. NO FURTHER ISSUES WERE IDENTIFIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OMNIPOD INSULIN PUMP PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 11200 L11873

Patients

Seq Age Sex Outcome Treatment
1 7 YR Other