FDA Adverse Event Injury Summary report: N

BOSTON SCIENTIFIC

MDR report key: 1321249 · Received February 17, 2009

Report

Report Number
MW5010036
Event Type
Injury
Date Received
February 17, 2009
Date of Event
February 16, 2009
Report Date
February 17, 2009
Manufacturer
BOSTON SCIENTIFIC
Product Code
KNT
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
CT, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

ON ATTEMPTING TO PLACE DRAINAGE TUBE, FROM BOSTON SCIENTIFIC, UNABLE TO USE CATHETER DUE TO INNER LUMEN BEING BLOCKED BY WHITE PLASTIC. THIS HAS HAPPENED NUMEROUS TIMES IN PAST TWO MONTHS. OF NOTE, THE CATHETER HAS A TIP PROTECTER THAT SLIDES OUT OF THE TIP PRIOR TO USE. THE PROTECTER WAS FOUND TO BE SNAPPED OFF IN THE LUMEN WHILE STILL IN THE PACKAGING LEAVING THE LUMEN BLOCKED. THIS HAS HAPPENED TO NUMEROUS CATHETERS SIZES #10 FR AND #8 FR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BOSTON SCIENTIFIC #10 FR ALL PURPOSE TT REGULAR FLEXIMA CATHETER SET KNT BOSTON SCIENTIFIC 9485168

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention